Get Started
248083296-1 Rev 3
3-5
Inlet Nozzle
Note:
The user is
responsible for
maintaining a nominal
flow rate of 100 ml/min +
5% through the sensor.
The sample source is attached using ¼" flare tubing to the Inlet Nozzle.
The user is responsible for maintaining a nominal flow rate of 100 ml/
min + 5% through the sensor to insure the accuracy of the data.
If there are any questions or problems during system setup, please
contact Lighthouse Worldwide Solutions’ Technical Support at
[email protected].
Figure 3-4 Example of an RLPC Installation
Sample Source
Valve
Valve
Liquid Flow
Meter
Pump
(Optional)
Sample Return
or to Drain
Inlet
Nozzle
Outlet
Nozzle
RLPC
sensor
Содержание RLPC 0.1mm
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counters 4 20mA Output Operating Manual...
Страница 8: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual t iv 248083296 1 Rev 3...
Страница 10: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual ii 248083296 1 Rev 3...
Страница 20: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 2 8 248083296 1 Rev 3...
Страница 29: ...Get Started 248083296 1 Rev 3 3 9 3 Wire System Figure 3 7 Three Wire Communication Diagram...
Страница 30: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 3 10 248083296 1 Rev 3...
Страница 50: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 5 4 248083296 1 Rev 3...
Страница 54: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual B 2 248083296 1 Rev 3...
Страница 59: ...00...
Страница 60: ......