Lighthouse REMOTE Liquid Particle Counter (4-20mA output) Operating Manual
2-2
248083296-1 Rev 3
imaged onto a photodiode. The photodiode converts the light into an
electrical current which is converted to voltage and amplified by the
electronics.
The result is a voltage pulse each time a particle crosses the LASER
beam. The width of the pulse is proportional to the time it takes the
particle to cross the LASER beam and the pulse’s amplitude is
proportional to the size of the particle.
The voltage pulses created by the particles are processed by additional
electronics to quantify the pulses by the size of each particle. The
quantities of the various size particles are processed and stored in the
sensor’s buffers or transferred via the 4-20mA interface.
The RLPC instruments were created for continuous, 24 hours per day, 7
days per week operation. The RLPC provides versatile mounting
options allowing installation where space is at a premium. The RLPC
integrates seamlessly with many large facility monitoring or
management systems.
Accessories
Several accessories can be ordered to tailor the instrument to specific
needs. The accessories are listed below.
Included:
•
Operating Manual on CD
•
2 each 10-foot flared sample tubes
•
Flow Cell Cleaning brush
•
Micro90
TM
Cleaning Solution
•
Power Supply
Optional:
•
Flowmeter (Teflon
®
PFA or aluminum)
•
Additional Sample Tubing
•
Cabling
•
REMOTE Mounting Plate
•
RLPC Stand with Flowmeter mounting
•
Flare nut kit with flaring tool and tubing cutter
•
Replacement Flare nuts
•
Replacement Flow Cell cleaning brushes
•
Additional Micro90
TM
Cleaning Solution
Содержание RLPC 0.1mm
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counters 4 20mA Output Operating Manual...
Страница 8: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual t iv 248083296 1 Rev 3...
Страница 10: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual ii 248083296 1 Rev 3...
Страница 20: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 2 8 248083296 1 Rev 3...
Страница 29: ...Get Started 248083296 1 Rev 3 3 9 3 Wire System Figure 3 7 Three Wire Communication Diagram...
Страница 30: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 3 10 248083296 1 Rev 3...
Страница 50: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 5 4 248083296 1 Rev 3...
Страница 54: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual B 2 248083296 1 Rev 3...
Страница 59: ...00...
Страница 60: ......