Programming
248083296-1 Rev 3
4-11
Type:
<sb>
Response:
<sb>STOP
GET CURRENT VERSION NUMBER
<gv>
Shows current version number of the device firmware.
Type:
<gv>
Response:
<gv> Version: 010
SET CHANNEL RANGES
Note:
Range 1 should
always be greater than or
equal to Range 2; Range 2
should never be greater
than Range 1.
When using the DIP switch
programming, both
channels are set to the
same range value.
<ra#>
-- sets upper range of channel 1 where # is the range
value
<rb#>
-- sets upper range of channel 2 where # is the range
value
Sets upper limit to counting range. The upper limit will be equal to
20mA, and the lower range will always be 0 counts = 4 mA. If the
counts exceed range limit, the counts will be set to the range limit. The
maximum range is 10,000,000.
Type:
<ra10000>
Response:
<ra10000>Range1= 10000.
Type:
<rb7500>
Response:
Содержание RLPC 0.1mm
Страница 1: ......
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions REMOTE Liquid Particle Counters 4 20mA Output Operating Manual...
Страница 8: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual t iv 248083296 1 Rev 3...
Страница 10: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual ii 248083296 1 Rev 3...
Страница 20: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 2 8 248083296 1 Rev 3...
Страница 29: ...Get Started 248083296 1 Rev 3 3 9 3 Wire System Figure 3 7 Three Wire Communication Diagram...
Страница 30: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 3 10 248083296 1 Rev 3...
Страница 50: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual 5 4 248083296 1 Rev 3...
Страница 54: ...Lighthouse REMOTE Liquid Particle Counter 4 20mA output Operating Manual B 2 248083296 1 Rev 3...
Страница 59: ...00...
Страница 60: ......