Operation - Mass Concentration Mode
5HY
5-17
UGR lists the Mass Concentrations for each individual particle size,
expressed in micrograms per cubic meter (ugr/m3).
Press BACK to return to the Configuration screen or press MAIN to
return to the MAIN screen.
ALARM
The user can enable alarming on specific channels as illustrated in
Figure 5-16.
Figure 5-16 Particle Alarm Configuration, 2 channels enabled
Note:
Alarming is only
applicable for AUTO and
MANUAL mode. It applies
only to Raw particle counts
even if the instrument is
displaying Normalized
data.
To enable the alarming for any channel, press the “
X
” next to that
channel. When a checkmark is displayed, that channel is enabled for
alarming. Press the checkmark to disable the alarming for that channel.
Alarm Threshold
Alarm thresholds are always compared to real-time particle counts,
even when the instrument is in Mass Concentration mode. The
instrument cannot be set to alarm on a specific level of micrograms per
cubic meter.
Содержание Handheld 3016 IAQ
Страница 1: ...OPERATING MANUAL Handheld3016IAQ P A R T I C L E C O U N T E R...
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions HANDHELD 3016 IAQ Particle Counter Gen E Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual ii 5HY...
Страница 14: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 1 2 5HY...
Страница 18: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 2 4 5HY...
Страница 78: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 4 50 5HY...
Страница 128: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual A 2 5HY...
Страница 130: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual B 2 5HY...
Страница 135: ...00...