Operation - Particle Mode
5HY
4-19
PARTICLE Display
Data on the instrument can be displayed in
Differential
(DIFF) or
Cumulative
(CUML) counts.
For example, the cumulative count for the 1
m channel is the sum of
that channel’s count + 2.5
m count + the 5
m count.
The differential count for the 1
m channel is the number of particles
between 1
m and 2.5
m.
The data displays on the MAIN screen according to whichever is
selected (DIFF or CUML).
The data format is either Raw (RAW) or Normalized (NORM).
Raw
data displays the actual number of particles counted.
Normalized
data
shows particle concentrations calculated from the raw data based on the
settings chosen (ft
3
or m
3
).
Volume of Air = Sample time (minutes) x FlowRate (CFM)
Normalized Data = Number of Particles/Volume of Air
Press BACK to return to the Configuration screen or press MAIN to
return to the MAIN screen.
ALARM
The user can enable alarming on specific channels as illustrated in
Figure 4-19.
Figure 4-19 Particle Alarm Configuration, 2 channels enabled
Threshold Button
Содержание Handheld 3016 IAQ
Страница 1: ...OPERATING MANUAL Handheld3016IAQ P A R T I C L E C O U N T E R...
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions HANDHELD 3016 IAQ Particle Counter Gen E Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual ii 5HY...
Страница 14: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 1 2 5HY...
Страница 18: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 2 4 5HY...
Страница 78: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 4 50 5HY...
Страница 128: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual A 2 5HY...
Страница 130: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual B 2 5HY...
Страница 135: ...00...