Maintenance
Procedures
This Chapter provides the user with the procedures to perform in
maintaining the HANDHELD instrument.
Safety
Before performing any of the maintenance tasks described in this
chapter, read Chapter 1 of this manual and become familiar with the
warnings and caution labels.
Calibration
To maintain optimum performance of this instrument, it should be
recalibrated annually by a Lighthouse Authorized Service Provider.
Cleaning
This procedure may be superceded by customer requirements;
however, not under any circumstances apply Acetone to the
HANDHELD instrument.
1.
To keep the interior of the instrument clean, remove the isokinetic
probe from the air inlet and install the protective plastic cap
supplied with the instrument.
2.
Moisten a lint-free cloth with isopropyl alcohol. The cloth should
be moist, not wet.
3.
Wipe down the exterior surfaces of the instrument.
Purge Test
1.
Connect the Purge filter to the sample inlet. The purge filter
should be a 0.1 micron, 0.1 CFM filter.
2.
Apply power to the instrument.
3.
Configure the unit via the touchscreen interface to sample for 30
minutes.
Содержание Handheld 3016 IAQ
Страница 1: ...OPERATING MANUAL Handheld3016IAQ P A R T I C L E C O U N T E R...
Страница 2: ...00...
Страница 3: ...Lighthouse Worldwide Solutions HANDHELD 3016 IAQ Particle Counter Gen E Operating Manual...
Страница 6: ......
Страница 12: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual ii 5HY...
Страница 14: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 1 2 5HY...
Страница 18: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 2 4 5HY...
Страница 78: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual 4 50 5HY...
Страница 128: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual A 2 5HY...
Страница 130: ...Lighthouse HANDHELD 3016 IAQ Gen E Operating Manual B 2 5HY...
Страница 135: ...00...