17
CHECKING
NOTE:
If opening the test strip vial for
the first time, write the discard date
(
4
months from today) on the vial label
and perform a control solution test fol-
lowing the instructions below.
4
.
Shake the control solution vial. Apply
1
drop of control solution to the centre of
the pink test square.
◆
The first time you open the control
solution vial, write the discard date on
the label in the space provided. The
discard date for control solution is
3
months from the date you first open
the vial.
◆
Check expiration date and discard
date. Do not use the control solution
after these dates.
◆
If the entire white pad surrounding the
pink test square is soaked, you have
applied too much Glucose Control
Solution.
CAUTION:
Applying too much control
solution may cause out–of–range results
and may make the meter dirty.