Odyssey® Sa Imager
loose and stick to something. If the instruments have been moved out of alignment, you may also
get errors if the BioStack Plate Stacker can’t pickup a plate or if it drops the plate incorrectly on the
Odyssey Sa Imager plate carrier.
In general, the software remedy for jams and other errors is to start the express run again. If you are
using express run templates, you can re-stack the plates, exclude the plates that have already been
successfully scanned and start the express run again. If you were running an express run that was
not saved as a tempate, you will need to define a new express run for the plates remaining in the
input stack.
In general, you will have the least difficulty with jams and other problems related to the BioStack
Plate Stacker if you use the recommended plates (Chapter 3) and don’t mix different types of plates
in the input stack.
Another category of errors that can occur are communication errors between the Odyssey Sa Imager
and the BioStack™ Plate Stacker. These errors are generally caused by unplugging the USB cable
and later plugging it back in. Communication is not always immediately established. The remedy is
to turn off both instruments, turn on the BioStack Plate Stacker, and then turn on the Odyssey Sa
Imager after the BioStack Plate Stacker has finished its startup routine.
Page 47