34) Press the eject button on the Odyssey Sa Imager to extend the plate carrier and verify there is
enough clearance between it and the BioStack Plate Stacker.
35) Slide the leveling and alignment fixture back in the plate gripper and verify that the BioStack
accessory is level. Adjust as needed.
BioStack X-Y Alignment
The BioStack Plate Stacker is now close to the final position, but some small movements will be
needed in the X-Y dimensions to ensure that microplates are placed squarely on the plate carrier.
36) With the leveling and alignment fixture still in the BioStack plate gripper, manually lower the
plate gripper until it is just above the Odyssey Sa Imager plate carrier. (Make sure the BioStack
stacker is off.)
37) Observe the position of the fixture relative to the four alignment pins on the plate carrier. When
the X-Y alignment is correct, the fixture will fit as shown below.