Odyssey® Sa Imager
27) Adjust the leveling feet on the BioStack stand until the bubble level indicates the plate gripper is
level. Adjust the feet only enough to level the instrument, while trying to maintain the same
height and clearance from the Odyssey Sa Imager plate carrier.
28) Move the plate gripper to its highest position and press the eject button on the Odyssey Sa
Imager to extend the plate carrier. Verify that the clearance between the BioStack linear bearing
and the Odyssey Sa plate carrier is still about 2 mm. Adjust the height and level the BioStack
Plate Stacker again as needed.
29) Push the Odyssey Sa Imager eject button to retract the plate carrier. Remove the leveling and
alignment fixture from the plate gripper.
Connecting the BioStack Plate Stacker to the
Locator
The BioStack stand has two large knobs on the front of the stand that tighten two alignment pins.
These alignment pins fit into the two holes on the locator clamp. The steps below can be used to
adjust the height of the alignment pins and to connect the BioStack stand to the locator clamp.
30) Loosen the large knobs on the bottom of the BioStack stand just enough so the alignment pins
move freely.
31) Observe the alignment pins below the BioStack stand. When they are adjusted properly the ball
on the end of the pins should be centered within the thickness of the locator clamp. If the pins
are too high or low, follow step 32 for height adjustment. Skip to step 33 if the pins are properly
adjusted.
Page 39