14
GA130001441_002_C0
– 09/2016
4.1.4
Standards
Conformity with respect to Low Voltage Directive 2014/35/EU
Conformity with respect to EMC Directive 2014/30/EU
Conformity with respect to RoHS Directive 2011/65/EU
International/national standards as well as specifications:
DIN EN 61010-1 (2011)
(Safety requirements for electrical equipment for measurement, control and
laboratory use).
DIN EN 61326-1 (2013)
(Electrical equipment for measurement, control and laboratory use
– EMC
requirements. Industrial interference immunity; electromagnetic emissions household
sector Class B).
4.2
Mains Connection
Voltage:
100
– 240 VAC
Frequency:
50/60 Hz
Fuses:
2 x T1.6A H
Power consumption
:
< 10 W
Current consumption:
0.2
– 0.1 A approx.
Overvoltage category:
II
Protection category:
1
Connection:
Cold-device plug IEC 320 C14
4.3
Measurement Channels
Number:
Two THERMOVAC measurement channels
One PENNINGVAC measurement channels
Connection:
6-way DIN socket for THERMOVAC
1-way HV socket (shielded) for PENNINGVAC
4.3.1
TM Measurement Channels
Measurement range:
5·10
-4
– 1000 mbar
Measurement inaccuracy:
± 20 % of measured value (10
-3
– 10
-2
mbar)
± 15 % of measured value (10
-2
– 10
2
mbar)
Cable length:
up to 100 m
Cable length alignment:
automatic
4.3.2
PM Measurement Channel
Measurement range:
1·10
-9
– 1·10
-2
mbar
Measurement inaccuracy:
± 30 % of measured value (10
-8
– 10
-4
mbar)
Cable length:
up to 100 m for the pressure range 1·10
-8
– 1·10
-2
mbar
4.3.3
Measurement
Display rate:
4 s
-1
Unit of measurement:
mbar, Pa, Torr