3.1.3 Instructions for Mechanical Installation
!
WARNING
Hazardous voltage may be present.
In order to avoiding fluid spills shorting voltages within the converter,
place the controller in a spill protected environment (for example
protected electronic cabinets).
If explosive flammable gases are present, place the converter in an
explosion-proof cabinet.
1. The converter shall be installed in a spill protected cabinet.
2. Assure that enough ventilation is provided in order to avoid exceeding the allowed ambient temperature and
humidity range:
Temperature range: 0
–40°C (32–104°F)
,
Humidity range: 30-85% R.H. (no condensation).
3. A wall-mountable extension cable (see
) is provided in order to panel mount the sensor
cable on the cabinet wall. Make a whole to the cabinet wall according to
The gasket ring delivered
with the extension cable can be used on the inside of the cabinet wall hence achieving an
IP-67
sealing.
4. For connecting the sensor to the wall-mountable connector on the cabinet the interconnect cable according to
can be used.
Step 1:
Align first the
Snap-Pin
side of the converter
with the DIN rail corner (picture
a
). If the rigid
side is aligned first (see picture
b
) than some of
the DIN rail parts of the converter housing might
be damaged.
Step 2:
Press in the converter on the spring-pin side
(see picture
c
).
Step 3:
Press in the fix side of the DIN rail recess of the
converter housing (see picture
d
)
Step 4:
Relax the converter and let the snap-pin fix it
(see picture
e
).
Figure 13: Din Rail mounting steps for converter LFC-1C-PC
(For dismounting follow the steps in reverse order.)
a
b
c
d
e
Snap Pin