Installation
3.1 Installation of Converter LFC-1C-PC
3.1.1 Overview and Preparation
Figure 11: Single channel converter LFC-1C-PC
PIN
Designation
Specification
Standard Configuration
Description
1
DC24V+
Voltage: 24 VDC ±10%
Current: 150 mA
Inrush current: 4.4 A 2ms
--
Supply voltage
2
DC24V-
3
FG
Field Ground
See note in installation
Section 3.1.2
Field ground
4
Analog Out +
4-20 mA or 0-20mA configurable
Load resistance < 600
Flow reading with 4-20 mA
Standard full scale flow range of each
sensor model.
Flow rate signal.
Update rate is 10ms.
5
Analog Out -
6
Digital Out1 +
Max. rating 30 V DC, 20 mA
Open collector
Configurable as Alarm High, Alarm Low,
Measurement Error, Volume Counter
Pulse, Volume Counter Alarm, Flow as
Frequency, Bubble Detect
- Parameter: Flow Alarm Low
- Setting: 12% of full scale
- Non contact type
Contact is made, when flow is below of 12% of
full scale.
7
Digital Out2 -
- Parameter: Measurement Error
- Non contact type
Contact is made, when measurement error is
occurred.
8
COM
--
Common Digital Out ground.
9
Digital In +
No-voltage contact or transistor oper.
Zero Adjustment
Is needed, if zero adjustment wants to be
triggered by PLC or when volume counter
function is used (integration of flow with volume
detection)
10
Digital In -
11
RS485 +
RS485
MODBUS
Protocol
Digital communication in a sensor array.
or configuration with service SW.
12
RS485 -
Table 9: Standard configuration of electrical interface (PLC and fieldbus)
Note 1: Min torque for connector screws is 0.5 Nm and max. is 0.6 Nm.
a
DIN Rail Mounting
Sensor Connectors
Interface Connector (12 Pin)
b
Re-Zero Button
High ADRS
LOW ADRS
4 Digit Display
c
LFC-1C-PC