User Manual for LEVIFLOW
®
LFSC-D Flowmeters
www.levitronix.com
PL-4507-00, Rev08, DCO# 20-024
22
6.
Fit the clamp hinge into the according sensor
body grove and close the clamp.
CAUTION
Take care to not squeeze the fingers during the
clamp-on process.
CAUTION
Assure that the tube is not squeezed and not axially
stretched during the clamp-on process.
7.
For a stable and repeatable measurement, it is
very important that the tube is properly clamped
into the square shape of the transducer area of
the sensor. Therefore, check that the sensor is
properly fitted into the square area and that the
tubing is not squeezed somewhere.
1.
After starting measurements with the system
with liquid temperature different from ambient,
or if the liquid temperature changes during
operation, full signal stability is reached after
about
30 min
.
Do not squeeze the tube
during the clamp-on
procedure!
A squeezed tube affects
flow measurement!