Description ODSL 30
Leuze electronic GmbH + Co. KG
ODSL 30
38
3.8
Advanced menu (for software versions V01.10 and newer)
In addition to the described functions, additional, new functions are available in the Advanced menu:
• Setting an offset/preset value to compensate for mounting tolerances
• Reduction in measurement time to as little as 30ms
• Changing the display resolution
Also available in the Advanced menu is the menu item
Applic. Param.
. This can be used to change
the measurement value output of the ODSL 30.
Activation of the advanced mode
Hold down the key during measurement operation for longer than 5s.
The
Advanced Menu?
display appears.
Press the or key to cancel activation of the Advanced menu.
Confirm
Yes
by pressing the button.
The
Advanced Menu is activated now
display appears briefly.
The menu item
Applic. Param.
is now also available in menu level 1.
3.8.1 Setting an offset/preset value - Compensating mounting tolerances
Deviations which occur during mounting of the ODSL 30 can be compensated for with the offset or preset
parameter:
• For Offset, a fixed value and a sign are specified.
• For Preset, a nominal measurement value is specified; a measurement is then performed using an
object located at the desired nominal distance.
NOTE
For information on querying the device software version, see see chapter 3.6.5.
NOTE
To protect against unintentional access, the Advanced menu is hidden from view by default and
must first be activated by the user.
ATTENTION
Please be certain to read the following notes before you activate the advanced mode and change
parameters in the menu item
Applic. Param.
.
ENTER
ENTER
ATTENTION
If the offset or preset results in negative measurement values, zero is output at the interface and
on the display.