16
R2931
TUV/ EU Power Cord
R2927
China/ AUS Power Cord
TRIVEX SYSTEM DISPOSABLES/ACCESSORIES
REF
DESCRIPTION
7205683
Autoclavable Storage and Sterilization Tray
7209513
Illuminator Inflow Tubing
7209514
TRIVEX , 4.5 mm Blade and Tubing
7209515
TRIVEX, 5.5 mm Blade and Tubing
7210115
Replacement Lamp
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The LeMaitre Vascular TRIVEX System is intended for use in the electromagnetic environment specified below. The customer or the user of the LeMaitre Vascular TRIVEX System
should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
The LeMaitre Vascular TRIVEX System uses RF energy only for its internal functions. Therefore, the
RF emissions are very low and are not likely to cause any interference in nearby electronic equip-
ment.
RF emissions
CISPR 11
Class A
The LeMaitre Vascular TRIVEX System is suitable for use in all establishments other than domestic
and those directly connected to the public low-voltage power supply network that supplies build-
ings used for domestic purposes
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Complies
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The LeMaitre Vascular TRIVEX System is intended for use in the electromagnetic environment specified below. The customer or the user of the LeMaitre Vascular TRIVEX System
should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Electrostatic (ESD) Discharge
IEC 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood, concrete or ceramic tile and the relative
humidity should be at least 30%.
Electrical fast transient/burst
IEC 61000-4-4
± 2kV for power supply lines
± 1kV for input/output lines
± 2kV for power supply lines
± 1kV for input/output lines
Mains power quality should be that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1kV differential mode
± 2kV common mode
± 1kV differential mode
± 2kV common mode
Mains power quality should be that of a typical commercial or
hospital environment
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
<95% dip in UT for 1/2 cycle
60% dip in UT for 5 cycles
30% dip in UT for 25 cycles
>95% dip in UT for 5 sec
<95% dip in UT for 1/2 cycle
60% dip in UT for 5 cycles
30% dip in UT for 25 cycles
>95% dip in UT for 5 sec
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the LeMaitre Vascular
TRIVEX System requires continued operation during power mains
interruptions, it is recommended that the LeMaitre Vascular
TRIVEX System be powered from an uninterruptible power supply
or a battery.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
The power frequency magnetic field should be at levels
characteristic of a typical location in a typical commercial or
hospital environment
NOTE: UT is the a.c. mains voltage prior to application of the test level.