79
Leica ASP300 S
Editing steps
• To edit a program step highlight the corresponding line and press the
respective table header.
• The number of cycles cannot be set to "
0
". To exclude the water or de-
tergent step, delete it out of the cleaning program.
The following characteristics of each step can be edited:
•
Reagent name
– only those reagents that are compatible with the re-
agent station being edited will be displayed.
•
Number of cycles
– the duration displayed is calculated based on the
number of cycles.
•
Retort temperature
– temperatures from 50 to 65 °C can be set for Ethanol
and Water and from 40 to 67 °C for Xylene.
7. Cleaning and Maintenance
The finish times of clean programs cannot be edited. Program dura-
tion is calculated automatically based on the number of cycles set.
The station 13-ext can be assigned exclusively to reagents from the
cleaning water group. It may only be used for cleaning programs.
Station 13-ext
• The designation "13-ext" is used for an additional station that can be used
in place of the standard station 13.
• Station no. 13 always contains cleaning water. For some applications it
is desirable to use a cleaning water/agent mixture between the cleaning
alcohol and the cleaning water.
• If a program contains the step "Station 13-ext", the operator is prompt-
ed to insert the reagent container "13-ext" at the required point in time
during the cleaning program, then to remove it once it is no longer used.