63
Leica ASP300 S
5.2.2 Replacing the paraffin
Draining the paraffin
• Connect the paraffin drain hose (
36
) to the
spout (
17
) of the paraffin drain at the front of
the instrument (
Fig. 55
) and into a catch tank.
Fig. 55
• Select the paraffin reservoir to be emptied and
the receptacle (External,
Fig. 56
).
Then, press
STATION DRAIN...
.
• This is followed by multiple security prompts
that you have to confirm individually in order
to continue.
17
36
From
FAVORITES
, call up the
SMART SCREEN
screen. (see previous chapter)
→
+
→
Fig. 56
5. Operation
• Use caution when handling melted
paraffin – molten paraffin is hot and
may cause burns.
• When connecting the hose, be sure
to press it onto the O-rings of the
drain opening as far as it will go.
• The paraffin drain hose must be
securely hooked into the external
receptacle and remain there through
the entire drainage process.
• When drainage is finished, the hose
is cleaned automatically using air.
• Do not remove the hose from the ex-
ternal receptacle until this cleaning
step is complete.