background image

Operation

38 

ARveo / Ref. 10 747 384 / Version 04

8.3.4  Setting the illumination field diameter

CAUTION
If the field diameter is greater than the field of view and the 
light intensity is too high, uncontrolled tissue heating may 
occur outside of the area visible through the microscope.

 

X

Do not set the light intensity too high.

Thanks to AutoIris, the field diameter is automatically adapted to 
the size of the field of view at the Leica M530 Optics carrier. 

 

X

To adjust the illumination field diameter manually, use rotary 
button (2). 
Automatic adjustment AutoIris is deactivated. 

 

X

To reactivate AutoIris press the Reset button (1).

2

1

If the illumination field diameter is blocked at a high light 
intensity in a high magnification setting, and cannot be 
adjusted automatically or manually, then the light intensity 
must be reduced in order to protect the tissue.

If the field diameter is locked in a small position and cannot 
be adjusted either automatically or manually, you can use 
an OR lamp to better illuminate a large field of view (small 
magnification position).

8.3.5  Adjusting the magnification (zoom)

You can adjust the magnification using a footswitch/handswitch or 
the "Magnification" adjustment bar on the "Main" menu screen of 
the control unit.

On the "Main" menu screen 

 

X

Press the 

 or 

 button on the bar for adjusting the 

magnification.

– or –

 

X

Press the magnification adjustment bar directly.
The magnification changes.

•  Clicking the 

 or 

 button changes the 

magnification value in increments of 1. Holding down 
the button with your finger changes the value in 
increments of 5.

•  You can adjust the magnification motor speed in the 

"Speed" menu.

•  These values can be saved individually for each user  

(see page 44).

WARNING
Danger to the patient due to failure of the magnification 
motor.

 

X

If the magnification motor fails, adjust the 
magnification manually.

Содержание ARveo

Страница 1: ...MEDICAL DIVISION User Manual 10 747 384 Version 04 ARveo...

Страница 2: ...Heerbrugg Tel 41 71 726 3333 Legal disclaimer All specifications are subject to change without notice The information provided by this manual is directly related to the operation of the equipment Medi...

Страница 3: ...ope on 34 8 2 Positioning the microscope 35 8 3 Adjusting the microscope 35 8 4 Transport position 41 8 5 Shutting down the surgical microscope 41 9 Control unit with touch panel 42 9 1 Menu structure...

Страница 4: ...e Indicates a potentially hazardous situation or improper use which if not avoided may result in appreciable material financial and environmental damage Information about use that helps the user to em...

Страница 5: ...d messages X Allocate responsibilities for commissioning operation and maintenance Monitor compliance with this X Only use the ARveo surgical microscope when it is free of defects X Inform your Leica...

Страница 6: ...ly hazardous optical infrared and UV radiation X Do not look at the operating lamp X Minimize exposure to eyes or skin X Use appropriate shielding WARNING Danger of burn injuries in otologic surgery X...

Страница 7: ...efore attaching the sterile drape check the counterweights for correct seating CAUTION Risk of infection X Leave sufficient space around the stand to ensure that the sterile drape does not come into c...

Страница 8: ...Safety notes 6 ARveo Ref 10 747 384 Version 04 3 5 Signs and labels 1 2 3 15 14 9 8 7 11 4 A 6 5 16 17 12 A 10 10 13...

Страница 9: ...Warning of hot surface 11 RISK GROUP 3 WARNING IR emitting from this product Do not look at operating lamp WARNING Possibly hazardous optical radiation emitted from this product Do not look at operati...

Страница 10: ...Safety notes 8 ARveo Ref 10 747 384 Version 04 1 2 3 5 2 4 2 2 1 Monitor weight label 2 Warning sign for squeezing hands or fingers 3 For installation only 4...

Страница 11: ...o stand 1 2 3 4 5 6 7 8 9 10 1 Arm system 2 Tension rod 3 Video monitor optional 4 Control unit with touch panel 5 Suspension device for footswitch 6 Illumination unit 7 Base 8 Interface panel 9 Lock...

Страница 12: ...17 9 Lock for installation only 11 Vertical arm 12 Interface panel 13 Footbrake 14 Handrail 15 Monitor arm 16 Camera control unit optional 17 Recording unit optional With its open architecture the AR...

Страница 13: ...visible light Leica HD C100 optional Interface for assistants either to the left and right side or to the back Main surgeon and back assistant interface 360 rotatable both Back assistant interface wit...

Страница 14: ...ier Leica M530 two directions of movement are balanced A and B 4 5 The essential performance of the ARveo is Providing sufficient light to the situs and to guarantee mechanical locking of the optics c...

Страница 15: ...rgical microscope 5 In the rotary joint 6 Brakes are operated via handle or footswitch if used The button of a handle footswitch with the assigned function Selected Brakes refer also to the chapter on...

Страница 16: ...225 mm Ev light intensity Iv brightness d distance from light source This means that less light is required to work with the microscope at a shorter distance than at a greater distance It is advisabl...

Страница 17: ...ing aid Leica SpeedSpot If Leica SpeedSpot is activated for the current user see page 49 the focussing aid is released when the brakes are released or when focussing Two convergent light beams meet ex...

Страница 18: ...agnification 8 Leica M530 Optics carrier 9 Switch for manual balancing of the A carriage 10 A carriage 11 Handle clamping lever 12 Switch for manual balancing of the B carriage 13 B carriage 14 Micros...

Страница 19: ...osystems Schweiz AG may be connected here AD F 1 and 2 are digital relay outputs that can switch 24 V 2 A 9 10 11 12 13 14 15 16 17 18 19 20 8 8 not in use 9 camera input optional 10 BNC IN 2x 11 BNC...

Страница 20: ...a Microsystems Schweiz AG may be connected here External Monitor Interface panel stand Ferrite 1 Ferrite 2 Ferrite 2 Ferrite 1 Requirements regarding EMC To reduce emission and ensure EMC compatibilit...

Страница 21: ...connecting the ARveo to an equipotential bonding device This is part of the customer s building installation Observe the requirements of EN 60601 1 8 6 7 7 Fuse box flap 9 10 8 Lamp inserts for main i...

Страница 22: ...brakes You can assign switches 1 2 3 and 5 of the handles individually for each user in the configuration menu In all presets key 4 releases all brakes This key cannot be configured For the joystick...

Страница 23: ...14 functions crosswise Footswitch 12 functions lengthwise Footswitch 16 functions lengthwise Footswitch 16 functions crosswise Footswitches can be assigned individually for each user in the configurat...

Страница 24: ...never pull it CAUTION Surgical microscope can move without warning X Always lock the footbrake when you are not moving the system NOTE Damage to the ARveo surgical microscope during transportation X N...

Страница 25: ...ever carry out the intraoperative AC BC balancing above the patient X Make sure that the optical accessories are clean and free of dust and dirt 7 3 Setting the binocular tube 7 3 1 Setting the interp...

Страница 26: ...he lens at working distance X Focus the microscope X Set the maximum magnification X Focus the microscope X Set the minimum magnification X Without looking into the eyepieces turn both eye lenses to 5...

Страница 27: ...the accessories or attempt to rebalance the microscope while it is above the field of operation X Balance the ARveo after re equipping it X Do not release the brakes when the instrument is in an unba...

Страница 28: ...on the control unit During the balancing procedure the push button flashes green and an acoustic signal sounds can be deactivated in the service menu 2 The following dialog window appears on the touc...

Страница 29: ...lancing push button on the control unit again During the balancing procedure the push button flashes yellow and an acoustic signal sounds can be deactivated in the service menu The following dialog wi...

Страница 30: ...irection 3 A direction Make sure no accessories collide with the microscope during manual balancing X Check the balancing X Press the All Brakes button on the handle Optics carrier tilting to the righ...

Страница 31: ...mize the position of the side assistant 7 6 3 Correcting the D balancing manually The internal weight 1 in the stand compensates for the weight of the surgical microscope and the installed accessories...

Страница 32: ...e to falling weight disk or cover X When changing the weight disk make sure that your feet are not beneath the weight disk or the cover X Detach the cover 3 from the axis 3 X Unscrew the hexagonal nut...

Страница 33: ...tion first swing the microscope away from the operating field X Never carry out the intraoperative AC BC balancing above the patient The ARveo can be positioned easily on the operating table and offer...

Страница 34: ...ries section CAUTION Risk of infection X Leave sufficient space around the stand to ensure that the sterile drape does not come into contact with non sterile components CAUTION Danger of injury due to...

Страница 35: ...d protective glass on the optics carrier so that the markings on the Leica M530 1 and on the protective glass 2 are aligned a b 1 2 X Insert the protective glass upwards into the bayonet mount in dire...

Страница 36: ...y X Adjust the field of view to match the operating field X Irrigate the wound frequently X Cover the exposed parts of the pinna with a moist surgical sponge X Connect the microscope to a grounded soc...

Страница 37: ...nel monitor a hand footswitch or handle On the touch panel monitor in the Main menu screen X Press the or button on the bar for adjusting the brightness of the illumination or X Press the brightness a...

Страница 38: ...rightness is reduced automatically It is advisable to begin with a low light output and increase the light intensity until an optimum level of illumination is achieved When shipped from the factory th...

Страница 39: ...rms you when the xenon lamp is losing luminosity and is no longer sufficient either for blue light FL400 application only or for white light all other applications We recommend that you keep a replace...

Страница 40: ...duced in order to protect the tissue If the field diameter is locked in a small position and cannot be adjusted either automatically or manually you can use an OR lamp to better illuminate a large fie...

Страница 41: ...e X When using lasers always set the working distance of the microscope to laser distance and lock the microscope in position X Do not adjust the rotary button for manual setting of the working distan...

Страница 42: ...Setting the working distance manually WARNING Danger of serious damage to tissue due to incorrect working distance X When using lasers always set the working distance of the microscope to laser dista...

Страница 43: ...l observers with zero diopters respectively with correct individual diopter settings This command can be as well given on the GUI and from the handle if defined 8 4 Transport position X Press the All...

Страница 44: ...sive substances These substances can scratch the surface and cause it to be become dull 9 1 Menu structure 1 2 3 4 5 1 Quick access to the screens Main Speed Menu DIC and Help 2 Status line 3 Display...

Страница 45: ...familiarize yourself with the assignment of the handles and the optional footswitch if used 9 2 2 Presets You can find a list of default users preset by Leica for the most common types of operation un...

Страница 46: ...he operational menu If you want to keep the current settings you can save them by clicking the Save button which appears as soon as the basic settings of the current users have been changed either for...

Страница 47: ...ain X Choose Save as current and enter the password PIN again If no password PIN was defined you can define a password PIN 4 10 characters X Press OK for re entering and confirmation If the re entered...

Страница 48: ...or each user in the user settings when All Brakes are released This function is deactivated in the factory default configuration 9 3 3 Setting the Speed start values For the selected user you can set...

Страница 49: ...croll forwards or backwards in the list by clicking the arrowheads X In the left selection field select the function group with the desired functions X You can scroll forwards or backwards in the list...

Страница 50: ...ft selection field select the function group with the desired functions You can scroll forwards or backwards in the list by clicking the arrowheads X Select the desired function X Click a free caption...

Страница 51: ...manuals 9 3 9 AutoFocus Settings AutoFocus is an optional function and can be ordered additionally AutoFocus is not available in all countries AutoFocus is not available with FL800 mode and FL400 mod...

Страница 52: ...entered 9 4 Menu Maintenance menu X Press the Menu button and select Maintenance The Maintenance menu offers the following screens Lamp History Check Switches Microscope Settings 9 4 1 Maintenance Lam...

Страница 53: ...le you want to test If the key you have pressed is functioning properly a green dot appears on it on the display The comment Tested appears in the caption field of the key 9 4 3 Maintenance Microscope...

Страница 54: ...gical microscope X Press the button for the topic desired Detailed information How to is displayed The Help button in the static menu bar provides access to the How To screens at all times 9 6 Menu Se...

Страница 55: ...ube var 0 180 T Type II Binocular tube var 30 150 T Type II L Inclined binocular tube T Type II Straight binocular tube T Type II Inclined binocular tube 45 Type II Eyepiece 10 Eyepiece 12 5 Eyepiece...

Страница 56: ...Footswitches Wireless Footswitch 14 functions Wireless Footswitch 12 functions Cart ITD for 31 and 55 Monitor See the corresponding instruction for use Do not use third party products without premissi...

Страница 57: ...on the following active ingredients aldehydes alcohols quaternary ammonium compounds Due to potential damage to the materials never use products based on halogen splitting compounds strong organic aci...

Страница 58: ...enance contract with them It is recommended to conclude a service contract with Leica Service Use only original spare parts for servicing After 18 months you will be reminded that the inspection is du...

Страница 59: ...row 4 is pointing to the left X Close the access door again The illumination push button item 2 section 6 2 lights up green Whenever you replace a bulb reset the bulb s hour meter to 0 by double click...

Страница 60: ...ol may also be used to clean the optics if heavy contamination such as fingerprints grease streaks etc is present X Dry off products except for optical components with a disposable cloth paper cloth D...

Страница 61: ...nsion 10446058 Protective glass multifocal lense 10448439 Protective glass 10448440 Cover sterilizable 10448431 Protective objective glass 10448296 Protective objective glass spare part package of 10...

Страница 62: ...ect X Check selection of the assistant see page 25 The image remains unfocused Eyepieces are not mounted correctly X Screw the eyepieces all the way on Diopters not set correctly X Perform dioptric co...

Страница 63: ...tall the video camera ARveo is not correctly balanced Position of accessory was changed after balancing X Balance out ARveo see page 25 X Perform intraoperative AC BC balancing see page 28 ARveo canno...

Страница 64: ...connecting cable has been disconnected or is defective X Check corresponding connection cable for proper seating and function X Contact your Leica representative Rear load too high The accessories be...

Страница 65: ...tics carrier Objective working distance 225 600 mm motorized multifocal lens continuously adjustable manual adjustment option status displayed in the display of the optics carrier Eyepieces Wide field...

Страница 66: ...Binocular tube Focal length Art No Type A f162 66 10447701 10446575 10448088 10446574 10446587 10446618 Type B f170 0 10446797 10448159 10448217 not recommended 14 2 3 Microscope carrier Rotation of...

Страница 67: ...the camera image size with respect to the field of view for the visual video camera Please be aware that the field of view is not fully covered by the documentation system 14 2 4 IGS Interface Compat...

Страница 68: ...lumination system and built in automatic lamp quick changer Control elements Pistol handle with 10 functions for magnification working distance All Brakes button releases 6 brakes side knob releases s...

Страница 69: ...471 EN 980 The Medical Division within Leica Microsystems Schweiz AG hold the management system certificates for the international standard ISO 13485 relating to quality management quality assurance a...

Страница 70: ...3 4 2 1 Leica M530 Optics carrier 2 Leica ULT530 or GLOW800 or Leica FL800 ULT 3 Leica ULT530 or GLOW800 or Leica FL800 ULT 4 Leica FL400 FL560 5 CaptiView Optional Side assistant Magnification Multip...

Страница 71: ...kg 10448088 S Binocular tube var 0 180 T Type II Not recommended vignetting 1 42 kg M Binoculartubeforbackassistant 10446797 S Binocular tube var 30 150 T Type II L Recommended 0 81 kg 10448088 S Bin...

Страница 72: ...eces main back sides 10448028 O Eyepiece 10x 2 eyepieces per binocular tube 0 10 kg 10448125 O Eyepiece 8 3x 0 10 kg 10443739 O Eyepiece 12 5x 0 10 kg 10448245 O Mouth switch 0 22 kg 10446058 O Protec...

Страница 73: ...Specifications ARveo Ref 10 747 384 Version 04 71 14 7 Dimensional drawings Dimensions in mm...

Страница 74: ...Specifications 72 ARveo Ref 10 747 384 Version 04 1200 max 1925 720 x 720 2685 1945 Dimensions in mm...

Страница 75: ...Specifications ARveo Ref 10 747 384 Version 04 73 1180 1785 430...

Страница 76: ...electromagnetic emissions The ARveo surgical microscope is intended for operation in an environment as specified below The customer or the user of the ARveo surgical microscope should make sure that i...

Страница 77: ...000 4 4 2 kV for power supply line 1 kV for input and output lines 2 kV for power supply line 1 kV for input and output lines Mains power quality should be that of a typical commercial or hospital env...

Страница 78: ...P in m 800 MHz up to 2 5 GHz d 2 4 P in m 0 01 0 24 0 24 0 24 0 1 0 8 0 8 0 8 1 2 4 2 4 2 4 10 8 0 8 0 8 0 100 24 0 24 0 24 0 For transmitters rated at a maximum output power not listed above the reco...

Страница 79: ...ing 5 Function check X Check the fiber optics cable connection to the optics carrier X Connect the power cable X Switch on the microscope X Switch on the illuminator at the control unit X Leave the il...

Страница 80: ...CH 9435Heerbrugg 2020 Printed 11 2020 Subjecttomodifications LEICAandtheLeicaLogoare registeredtrademarksofLeicaMicrosystemsIRGmbH Leica Microsystems Schweiz AG Max Schmidheiny Strasse 201 CH 9435 He...

Отзывы: