Preparation before surgery
ARveo / Ref. 10 747 384 / Version 04 29
Optics carrier tilting back/forwards in the B-position
3
X
Move the B axis with switch (3) until the optics carrier is
balanced.
Optics carrier tilting back
move B axis forwards
Optics carrier tilting forwards move B axis back
If the microscope cannot be balanced manually, the weight
and/or possition of the accessories is probably out of the
allowed range.
X
Reduce or increase the weight to the permitted range
and/or optimize the position of the side assistant.
7.6.3 Correcting the D-balancing manually
The internal weight (1) in the stand compensates for the weight of
the surgical microscope and the installed accessories.
It may be necessary to correct the D-balancing after fitting
a sterile drape on the microscope
D
D
1
X
Correct the D-balancing of the stand with the "if Scope is
Rising" and "if Scope is Falling" keys on the "Main" screen of the
control unit.
Microscope is too heavy
touch "if Scope is Falling" key
Microscope is too light
ouch "if Scope is Rising" key
To balance the D axis when using accessories with different
weights, the number of D axis weight disks can be adapted
accordingly (see below).