Preparation before surgery
28
ARveo / Ref. 10 747 384 / Version 04
If the optics carrier is not oriented correctly, the following dialog
window opens:
X
Correct the orientation of the optics carrier (B-position).
X
Press the autobalancing push-button.
Autobalancing re-starts.
7.6.2 Manual balancing of the ARveo
For manual balancing, the axes can be moved manually using
switches (1), (2) and (3).
3
1
2
1 C direction
2 B direction
3 A direction
Make sure no accessories collide with the microscope during
manual balancing.
X
Check the balancing.
X
Press the "All Brakes" button on the handle.
Optics carrier tilting to the right/left
1
X
Move the C axis with switch (1) until the optics carrier is
balanced.
Optics carrier tilting to the right
move switch to the left
Optics carrier tilting to the left
move switch to the right
Optics carrier tilting back/forwards
2
X
Move the A axis with switch (2) until the optics carrier is
balanced.
Optics carrier tilting back
move A axis forwards
Optics carrier tilting forwards move A axis back