151
HistoCore SPECTRA CV
Malfunctions and Troubleshooting
8
1
Fig. 124
3. If this is the case, push the sledge
of the transfer station back into the
HistoCore SPECTRA ST manually
and remove the rack
from the
holder
1
2
3
4
Fig. 125
4. Following this, remove the rack from the instrument and store it safely.
5. After the instrument fault has been corrected, insert the rack in the load drawer of the
HistoCore SPECTRA CV to start processing.
Содержание 14051454200