115
HistoCore SPECTRA CV
Cleaning and Maintenance
7
Note
•
Carefully attach the suction cups with a bent tweezers to avoid damage to the new suction cups.
• Make sure that the suction cups fit correctly. Coverglass cannot be picked up if the suction cups are
fastened incorrectly.
• We recommend having a rack filled with empty slides coverslipped for a function check.
2
1
1
3
Fig. 92
7.2.12 Cleaning the waste tray
Warning
Wear cut-resistant safety gloves
P. 18 – 3.1 Standard delivery – packing list)
for emptying the
waste tray!
For handling, removing and re-inserting the waste tray, see
P. 86 – 6.3.4 Emptying the waste tray)
.
» Soak a lint-free cloth with xylene and remove any mounting medium residue that may be present.
Note
Dispose of glass shards, splinters and dust in accordance with laboratory specifications.