Aperio AT2 DX User’s Guide, Revision A © Leica Biosystems Imaging, Inc. 2019
44
Chapter 7: Quality Control and Maintenance
Cleaning the AutoLoader Racks
Inspect racks for damage and for accumulated glass dust or
accumulated mounting media. All grooves within the rack (sides
and bottom) must be
clean and smooth. Clean the racks as required with compressed air and/or cleaning
solvent. Damaged
racks should be replaced immediately.
Clean racks frequently.
Recalibrate the AutoLoader Slide-in-Rack Sensor
Recalibrating the AutoLoader slide-in-rack sensor prevents scanning problems by ensuring that the slide moving from the
AutoLoader to the scanning stage is precisely positioned. This procedure should be done every six months. (This will be
done as part of your annual maintenance visit by Leica Biosystems Technical Services.)
First you will restart the scanner Controller service. This initializes the scanner and puts the stage and AutoLoader in the
home position.
1.
Turn the scanner off and back on: Press the main power switch on the back of the scanner to the Off position then
press it to the On position.
2.
Click the Service Manager icon
in the Windows task bar, select
Restart Services
, and then select
Restart
Controller Service
.
Now you will recalibrate the AutoLoader slide-in-rack sensor:
1.
Open the AT2 DX covers to gain access to the inside of the scanner.
2.
Locate the slide-in-rack sensor amplifier which is
to the right of the AutoLoader cover.