Doc No. FM0600 issue I Page 1
169MHz Telecare Devices
Sensor Controller
_______________________________________________________________________________________________________________
•
Compatible with Reach at-home alarms, Advent xt warden call,
Altec Response and Touchsafe Pro Nursecall systems
•
Can be used for bed or chair occupancy monitoring
•
Fully configurable with text display menus
•
Real time clock with auto-BST adjustment
•
Supplied with mains adapter & rechargeable battery
•
Dimensions: 190mm x 100mm x 32mm (HxWxD)
•
Weight: 285 grams (inc battery)
•
Sensor Controller with standard bed sensor: Product Code ZXT473
•
Sensor Controller with standard chair sensor: Product Code ZXT474
What is the Sensor Controller used for?
The sensor controller is normally used for bed or chair occupancy monitoring; if a person gets into bed/sits in their chair, then gets up
again and does not return within a preset period an alarm call will be transmitted. Alternatively, the sensor controller can be used with a
floor pressure mat or a door contact to raise an instant alarm.
How does the Sensor Controller work?
The sensor controller is normally supplied with a standard bed or chair sensor as a complete package. The controller and other sensor
options are also available separately if custom packages are required.
The sensor controller must be installed with a Reach at-home alarm, an Advent xt warden call system, an Altec Response local carer
alarm unit or a Touchsafe Pro Nursecall system.
The sensor controller has a real time clock with automatic British Summer Time adjustments (i.e. the +/- 1 hour clock changes are made
automatically on the correct day every March and October).
With a standard bed
or chair sensor an alarm can be raised either
instantly
a person leaves their bed/chair or after a 1 to 99 minute
programmable
absence period
if the person does not return to their bed/chair. If the person does not return and an alarm is raised but
they are OK, then an optional 1 to 99 minute programmable
reset period
can be set. If the person is not detected back in their bed/chair
at the end of this period, then an alarm will be raised again.
Monitoring starts when a person has been in their bed/chair for 30 seconds (default
logged in bed time
). Continuous monitoring is
possible if the controller is set to be
on all the time
. Alternatively, up to
3
active periods
can be programmed to repeat every day. There
is also the option to raise an alarm if a person is
not in their bed/chair
after a 0 to 999 minute programmable delay from the start of the
monitoring period.
Monitoring will end at the end of each monitoring period unless the
still in bed/chair
option is set. This is a 0 to 999 minute
programmable period – if the person is still in their bed/chair at the end of this period an alarm will be raised.
If the person is going to be away from home for a few days, the Sensor Controller has an
away mode
feature which suspends monitoring
to prevent false alarms. When the person returns home the unit must be switched back into
home mode
and normal monitoring will
resume.
The sensor controller has a
Carer Call
button to activate an alarm call at any time. An optional Pear Push lead is available to activate a
carer call.
With a Magic Stick or SensAlert bed/chair sensor an alarm can be raised either
instantly
a person leaves their bed/chair or after a
programmable 0 to 999 second
countdown
delay. The alarm can be cancelled during the countdown by pressing the
key.
With a pressure mat or door contact an alarm can be raised either
instantly
a person steps on the mat or opens the monitored door or
after a programmable 0 to 999 second
countdown
delay. The alarm can be cancelled during the countdown by pressing the
key.
An RJ45 double adapter is used to connect 2 sensors per controller -
Note: both sensors must be the same type
.
Tynetec and Aidcall are business units of Legrand Electric Ltd
Unit 10 Cowley Road, Blyth Riverside Business Park, Blyth
Northumberland, NE24 5TF. Tel:
01670 352 371
Fax:
01670 362 807
Email:
Web
: