Instruction Manual
ZD200-GSM-REVA
21
9.
Connect the PROBUS-CF01 BNC male output (the probe
end) to the precision 50Ω termination BNC input. The
probe should remain powered.
10.
Install the straight tips on the ZD200 inputs.
11.
Connect the signal generator output to the ZD200
inputs using the BNC to mini-grabber cable.
12.
Read the voltage from the DMM and record this value
on the test data sheet.
13.
Record the calculated error to two decimal places
(±0.xx%) as "Gain Error" in the test record.
14.
Verify that the error is less than ±1.0 %.
This completes the Performance Verification of the ZD200.
Complete and file the Test Record, as required to support your
internal calibration procedure.
Apply suitable calibration label to the ZD200 housing as
required.