4
–
7
Note: The default Gateway is assigned as "172.25.0.1" unless
your network has this Gateway available, you must ensure the
computer and the LSA1000 are on the same subnet.
Changing IP Address
Once communication between the LSA1000 and host PC has
been verified, the IP address can be easily reset to any address
using the
Scope Explorer
remote control software tool provided
in the Getting Started CD-ROM. To change the IP address,
initiate the following remote commands from
ScopeExplorer
’s
Terminal. These are described in the
Remote Control Manual
(see Chapter 5 of the present manual).
➢
*IDN?
: Retrieves the unit’s name, serial number, and
software revision level.
➢
CONET?
: Returns the current IP address configuration
including the mask and gateway. Default is IP, "172.25.1.2",
Mask, "255.255.0.0", Gateway, "172.25.0.1".
➢
CONET
IP,”172.25.1.5”,MASK,”255.255.0.0”,GATEWAY,”172.25.0.1”
:
Sets the IP address. This example changes the unit’s IP
address to 172.25.1.5.
Once the IP address is changed, the unit will no longer respond to
the original IP.
If the network settings are unknown or accidentally set to invalid
values, the factory default network settings can be recovered by
following the procedure below.
Restoring Default Network Settings
The factory default network settings can be
restored by powering-up the unit with the Ethernet reset jumper
closed. (The reset jumper is accessible through the small hole next
to the Ethernet connector).
Содержание LSA1000
Страница 2: ...Operator s Manual LeCroy LSA1000 Signalyst Revision B October 1998 ...
Страница 6: ...iii Contents ...
Страница 9: ...v Foreword ...
Страница 14: ...1 1 Read This First 1 ...
Страница 28: ......
Страница 29: ...2 1 2 About the LSA1000 ...
Страница 34: ...3 1 3 Installation and Safety ...
Страница 42: ...4 8 Connecting to PC or Network Example of responses for IDN and CONET queries ...
Страница 43: ...4 1 4 Connecting to PC or Network ...
Страница 55: ...5 12 LSA1000 Software Tools Remote LSA display of LSA1000 s External Reference Out Waveform ...
Страница 63: ...5 20 LSA1000 Software Tools ...
Страница 66: ...5 1 5 LSA1000 Software Tools ...
Страница 72: ......
Страница 73: ...A 1 A Appendix A Specifications ...
Страница 83: ......
Страница 84: ...B 1 Appendix B NET_CON Source Code B ...
Страница 96: ......
Страница 97: ...C 1 C Appendix C Parameter Measurement ...
Страница 104: ...D 1 D Appendix D Program Examples ...
Страница 107: ...E 1 E Appendix E Suggestion Forms ...
Страница 108: ...Index ...
Страница 109: ......
Страница 114: ...Index ...