C
–
3
Parameter Measurement
Determining Differential
Time Measurements
Figure C
–2
To avoid these bias effects, the instrument uses cyclic
parameters, including
crms
and
cmean
, that restrict the
calculation to an integer number of cycles.
The LSA1000 enables accurate differential time measurements
between two traces
— for example, propagation, setup and hold
delays (
Fig. C
–3
).
Parameters such as
c2d
require the transition polarity of the
clock and data signals to be specified.
RIGHT CURSOR
LEFT CURSOR
TWO FULL PERIODS:
= 2
cycles
first
last
50 %
delay
PERIOD
PERIOD
mean, median, rms, sdev
computed on interval periods
width
width
width
freq
period
= 1/
duty width/period
=
area, points, data
computed between cursors
TRIGGER
POINT
(Mesial)
Содержание LSA1000
Страница 2: ...Operator s Manual LeCroy LSA1000 Signalyst Revision B October 1998 ...
Страница 6: ...iii Contents ...
Страница 9: ...v Foreword ...
Страница 14: ...1 1 Read This First 1 ...
Страница 28: ......
Страница 29: ...2 1 2 About the LSA1000 ...
Страница 34: ...3 1 3 Installation and Safety ...
Страница 42: ...4 8 Connecting to PC or Network Example of responses for IDN and CONET queries ...
Страница 43: ...4 1 4 Connecting to PC or Network ...
Страница 55: ...5 12 LSA1000 Software Tools Remote LSA display of LSA1000 s External Reference Out Waveform ...
Страница 63: ...5 20 LSA1000 Software Tools ...
Страница 66: ...5 1 5 LSA1000 Software Tools ...
Страница 72: ......
Страница 73: ...A 1 A Appendix A Specifications ...
Страница 83: ......
Страница 84: ...B 1 Appendix B NET_CON Source Code B ...
Страница 96: ......
Страница 97: ...C 1 C Appendix C Parameter Measurement ...
Страница 104: ...D 1 D Appendix D Program Examples ...
Страница 107: ...E 1 E Appendix E Suggestion Forms ...
Страница 108: ...Index ...
Страница 109: ......
Страница 114: ...Index ...