Technical Note No. 037 

 

Procedure  for  Programming  Model  211  Ozone  Monitor  and  Model  405  nm 
NO

2

/NO/NO

x

 Monitor 

 
Date:  

29 April 2014

 

 
Author:  

Craig Williford

 

 

 

Requirements 

Programming the Model 405 and 211 monitors requires a computer running the "Analog 
Devices Serial Downloader" program to download the firmware from the computer to the 
monitor, and the 

straight through

 rs232 serial cable provided by 2B Technologies.  The 

computer must have a serial port or a USB port with a USB to Serial converter cable to 
connect to the rs232 serial cable provided by 2B Technologies.   
 

Procedure 

 

I.  Install Windows Serial Downloader Application

 

1.  If not already installed, please install the WSD.exe application provided by 2B 

Technologies.  

 

2. 

In the “Configuration” menu on the WSD, set the Port number to the COM port 
that corresponds to the one your serial connector is connected to.  Normally 
this is COM1 unless you do not have a serial port on your computer and you are 
using a serial to USB converter.  If using the serial to USB converter, check in 
Windows Device Manager for the correct port number corresponding to the 
converter.  

 

 

II.  Record all stored parameters

 

1.  Enter the main menu and write down or record calibration parameters (CF, SF, 

NO2 and NO slope, NO2 and NO zero), averaging frequency.  This is 
imperative as you will need to re-enter these parameters after programming the 
new firmware.

 

 

III.  Removal of top cover  

1.  Remove the six screws on the tops of the sides securing the top cover on. 
2.  Remove the unscrewed lid.  
 

IV.  Programming instrument 

1.  Plug in 12V power source for instrument. 