Adjustment Procedure
HFP1500-OM-E Rev E
8–3
Figure 8-1. Output Zero Voltage Adjustment Setup
3. Select the channel to which the probe and ProBus extender is connected. Set
OFF-
SET
on the oscilloscope to zero as indicated on the on-screen display.
4. Set the DMM to read DC Volt on the most sensitive range.
5. Verify that the probe inputs are not connected to any signal.
6. Adjust
OFFSET ZERO
on the board until the DMM reads 0 V ±100
µ
V. Refer to figure
8-2 for adjustment location.
7. Leave the setup connections in place for the next step.
Figure 8-2a. Adjustment Locations S/N 1000 and higher.
HF
P
25
00
Precision 50
Ω
Terminator
ProBus Extender Cable
HFP1500 probe
DMM
LeCroy
Oscilloscope
OUTPUT ZERO
OFFSET RANGE
OUTPUT ZERO
OFFSET RANGE
Содержание HFP1500
Страница 1: ...LeCroy HFP1500 High Frequency Probe Instruction Manual Revision F February 2008 ...
Страница 4: ...iv HFP1500 OM E Rev E BLANK PAGE ...
Страница 18: ...HFP1500 High Frequency Probe 3 8 HFP1500 OM E Rev E BLANK PAGE ...
Страница 22: ...HFP1500 High Frequency Probe 4 4 HFP1500 OM E Rev E BLANK PAGE ...
Страница 26: ...HFP1500 High Frequency Probe 5 4 HFP1500 OM E Rev E BLANK PAGE ...