Operation
HFP1500-OM-E Rev E
4–3
Note:
Probe offset is controlled with the channel OFFSET knob in oscilloscopes with
software version 8.7.0 or higher. The current offset is displayed above the grati-
cule for a few seconds after a change has been made.
USE WITH NON-LECROY INSTRUMENTS
The HFP1500 can be used with other instruments with the optional ADPPS power supply. The out-
put of the ADPPS must be terminated into 50
Ω
.
AutoColor ID feature and probe offset will be disabled when the probe is used with an ADPPS
adapter.
When used with an ADPPS adapter, the probe’s performance will be limited to 1 GHz and offset is
only possible by using the oscilloscope’s internal offset control.
Because the probe offset is not being used, the linear operating range is limited to ±8 Volt.
USE WITH OLDER LECROY INSTRUMENTS
When used with LeCroy instruments with software version lower than 8.7.0, the oscilloscope will
provide the correct scale factor but no AutoColor ID. Probe offset is also disabled.
Because the probe offset is not being used, the linear operating range is limited to ±8 Volt.
With V/div settings greater than 2 V/div, it is possible to display clipped waveforms on screen.
###