HFP1500 High Frequency Probe
7–4
HFP1500-OM-E Rev E
Figure 7-1. Output Zero Voltage Test Setup
3. Set the
OFFSET
on the oscilloscope to zero, as indicated by on-screen display.
4. Set the DMM to read DC Volt on the most sensitive range.
5. Record the voltage measured on the DMM to 10 µV resolution as ’Output Zero Voltage’ in
the Test record.
6. Check that the voltage indicated by the DMM is between ±800
µ
V.
7. Disconnect the DMM from the precision 50
Ω
terminator. Leave the remaining setup in
place for the next step.
H
F
P
250
0
ProBus Extender Cable
Precision 50
Ω
Terminator
HFP1500 Probe
DMM
LeCroy
Oscilloscope