12. Vector Display
12-2
12.2
Vector and Scale Settings
To configure vector and scale settings, press F•1 INTEN/SCALE in the vector menu. You can
configure the I and Q axes display, the vector intensity, and the scale intensity and colors.
VECTOR
→
F•1 INTEN/SCALE
→
Figure 12-2 INTEN/SCALE menu
12.2.1
Setting the Vector Intensity
To set the vector intensity, follow the procedure below. Regardless of this setting, in the
multi-screen display, scales are displayed using the intensity that MULTI VEC is set to. The
VECTOR INTEN value set using MULTI VEC and the WFM INTEN value set using MULTI
WFM are the same.
If you press F•D, the intensity will be reset to its default value of 0.
This setting is valid when DISPLAY is set to VECTOR.
Reference: MULTI
VEC
→
Section 16.2, “Setting Each Measurement Mode”
DISPLAY
→
Section 12.6, “Switching between the Vector, 5 Bar, and Phase Difference
Displays”
Procedure
VECTOR
→
F•1 INTEN/SCALE
→
F•1 VECTOR INTEN
Settings
Selectable range: –128 to 127 (The default value is 0.)
12.2.2
Setting the Scale Intensity
To set the scale intensity, follow the procedure below. Regardless of this setting, in the
multi-screen display, scales are displayed using the intensity that MULTI VEC is set to. The
SCALE INTEN value set using MULTI VEC and the SCALE INTEN value set using MULTI
WFM are the same.
If you press F•D, the scale intensity will be reset to its default value of 4.
Reference: MULTI
VEC
→
Section 16.2, “Setting Each Measurement Mode”
Procedure
VECTOR
→
F•1 INTEN/SCALE
→
F•2 SCALE INTEN
Settings
Selectable range: –8 to 7 (The default setting is 4.)