- 2 -

CONNECTING  SEVERAL  CELLS  IN  PARALLEL

:  Connect  several  cells  in  parallel  by  using  -  if 

necessary - a watertight junction box with terminal box. The cell connection extension cables must 

be shielded, led individually into their piping or conduit and laid as far as possible from the power 
cables (in case of 4-wire connections, use cables with 4x1 mm

2

 minimum cross-section). 

WELDING

: Avoid welding with the load cells already installed. If this cannot be avoided, place the 

welder ground clamp close to the required welding point to prevent sending current through the load 

cell body. 

WINDY  CONDITIONS  -  KNOCKS  -  VIBRATIONS

:  The  use  of  weigh  modules  is  strongly 

recommended for all load cells to compensate for misalignment of the support surfaces. The system 

designer  must  ensure  that  the  plant  is  protected  against  lateral  shifting  and  tipping  relating  to: 

shocks and vibration; windy conditions; seismic conditions in the installation setting; stability of the 

support structure. 

EARTHING THE WEIGHED STRUCTURE

: By means of a copper wire with suitable cross-section, 

connect the cell upper support plate with the lower support plate, then connect all the lower plates to 

a single earthing system. Electrostatic charges accumulated because of the product rubbing against 

the pipes and the weighed container walls are discharged to the ground without going through or 

damaging  the  load  cells.  Failure  to  implement  a  proper  earthing  system  might  not  affect  the 

operation of the weighing system; this, however, does not rule out the possibility that the cells and 

connected instrument may become damaged in the future. It is forbidden to ensure earthing system 

continuity by using metal parts contained in the weighed structure. 

FAILURE TO FOLLOW THE INSTALLATION RECOMMENDATIONS WILL BE CONSIDERED 

A MISUSE OF THE EQUIPMENT 

OK

OK

NO

NO

NO

OK

Содержание W200-MU

Страница 1: ...Base 2014 30 EU EN55022 2010 EN61000 6 2 2005 EN61000 6 4 2007 SYSTEM IDENTIFICATION Microlectra bv www microlectra nl info microlectra nl Microlectra bv Augustapolder 12 2992 SR Barendrecht The Neth...

Страница 2: ...er No guarantee against misuse Batteries Laumas provides 1 year guarantee from the date of delivery note against material defects or battery manufacturing faults Disposal of Waste Equipment by Users i...

Страница 3: ...EM PARAMETERS 12 THEORETICAL CALIBRATION 12 MAXIMUM CAPACITY 13 TARE WEIGHT ZERO SETTING 13 ZERO VALUE MANUAL ENTRY 13 REAL CALIBRATION WITH SAMPLE WEIGHTS 14 FILTER ON THE WEIGHT 15 ANTI PEAK 15 ZERO...

Страница 4: ...S485 CONNECTION 30 RS232 CONNECTION 30 TEST 31 DATE AND TIME SETTING 31 INFO MENU 31 SETPOINT PROGRAMMING 32 ALARMS 32 PRINTING EXAMPLES 34 RESERVED FOR THE INSTALLER 35 MENU LOCKING 35 MENU UNLOCKING...

Страница 5: ...emote control switch coils Avoid inverters in the instrument panel if inevitable use special filters for the inverters and separate them with sheet metal partitions The panel installer must provide el...

Страница 6: ...l shifting and tipping relating to shocks and vibration windy conditions seismic conditions in the installation setting stability of the support structure EARTHING THE WEIGHED STRUCTURE By means of a...

Страница 7: ...itation wire must be equal or similar to the one indicated in the load cell data sheet input resistance The insulation value between the shield and any other cell wire and between any other cell wire...

Страница 8: ...ensions 180x130x77 mm four fixing holes 4 mm centre distance 163x113 mm Two serial ports RS485 and RS232 for connection to PC PLC up to 32 instruments max 99 with line repeaters by ASCII Laumas or Mod...

Страница 9: ...conversions second DISPLAY RANGE 999999 NO OF DECIMALS DISPLAY INCREMENTS 0 4 x 1 x 2 x 5 x 10 x 20 x 50 x 100 DIGITAL FILTER READINGS PER SECOND 10 levels 5 300 Hz RELAY OUTPUTS N 5 max 115 VAC 150 m...

Страница 10: ...115 230 VAC optional version 12 INPUT COMMON VDC 0 V N NEUTRAL 115 230 VAC optional version 13 GROUND 115 230 VAC optional version 14 INPUT No 3 VDC min 5 V max 24 V otherwise ANALOG OUTPUT 0 20 o 4 2...

Страница 11: ...the welding side join the red points using a stiff wire RXD 19 1 13 2 3 4 5 6 16 1415 1718 TXD 12 24 7 8 11 10 9 21 20 22 23 IN 1 IN 2 IN COMMON INPUTS supply 5 24 VDC 0 20 4 20 mA COMMON 0 10 VDC 1...

Страница 12: ...splay press and hold down MENU and keys at the same time press MENU immediately followed by KEY Short press Long press 3 s Into menus Semi automatic zero Tare resetting Cancel or return to previous me...

Страница 13: ...9 MENU MAP Into menus changes are applied right after pressing the ENTER key no further confirmation is required SETPOINT SYSTEM PARAMETERS 3 s optional...

Страница 14: ...10 3 s optional optional...

Страница 15: ...tion REAL CALIBRATION WITH SAMPLE WEIGHTS If the instrument HAS NOT BEEN CALIBRATED missing plant system identification tag proceed with calibration If load cells data are unknown follow the procedure...

Страница 16: ...d on the load cells It s possible to set a value between 0 50000 and 7 00000 mV V Example of 4 cell system with sensitivity 2 00100 2 00150 2 00200 2 00250 enter 2 00175 calculated as 2 00100 2 00150...

Страница 17: ...compensate zero variations due to the presence of product residues Procedure Confirm the message by pressing ENTER The weight value to be set to zero is displayed In this phase all of the LEDs are fla...

Страница 18: ...the system must read zero If this does not happen it means that there is a mechanical problem affecting the system linearity WARNING identify and correct any mechanical problems before repeating the...

Страница 19: ...ntil an optimum result is achieved The filter enables to stabilise a weight as long as its variations are smaller than the corresponding response time It is necessary to set this filter according to t...

Страница 20: ...on the weight value is lower than the value set in this parameter and does not exceed the value the weight is reset To disable this function set 0 ZERO TRACKING from 1 to 5 default When the weight va...

Страница 21: ...e inputs is set to mode see section OUTPUTS AND INPUTS CONFIGURATION when the input is closed the value will be displayed modified according to the coefficient when the input is opened the standard we...

Страница 22: ...LE value divided by the conversion coefficient from kg to the new unit of measure Example The 4 load cells of 1000 kg are placed under a scale for olive oil which has a specific gravity of 0 916 kg l...

Страница 23: ...ht the contact will switch on the basis of net weight If the net function is not active the contact will switch on the basis of gross weight If net functions are disabled see section NET FUNCTIONS by...

Страница 24: ...displayed based on the set coefficient see setting of the units of measure and coefficient otherwise the weight is displayed when the input is closed the data are sent for printing if in the communic...

Страница 25: ...n can be repeated many times by the operator to allow the loading of several products Example Put the box on the scale the display shows the box weight press TARE the display shows the net weight to z...

Страница 26: ...ed keep pressed to display the gross weight When the key is released the net weight will be displayed again IF A SEMI AUTOMATIC TARE NET IS ENTERED IT IS NOT POSSIBLE TO ACCESS THE ENTER PRESET TARE F...

Страница 27: ...For the output 10 V and 5 V the soldered jumper SW1 must be closed open the instrument releasing with a screwdriver the locking tabs that hold together the two sides of the case locate on the printed...

Страница 28: ...ust set the parameter to 20 1 to get 20 0 on the PLC or tester Minimum and maximum values which can be set for zero and full scale corrections ANALOG OUTPUT TYPE Minimum Maximum 0 10 V 0 150 10 200 0...

Страница 29: ...ts settings set continuous weight transmission protocol to RIP6100 RIP675 RIP6125C series remote displays the remote display shows the net weight or gross weight according to its settings set continuo...

Страница 30: ...200 Hz with minimum baud rate 38400 baud 300 Hz with minimum baud rate 38400 baud delay in milliseconds which elapses before the instrument replies from 0 to 200 ms default 0 no parity default even p...

Страница 31: ...is necessary to connect the above mentioned resistors DIRECT CONNECTION BETWEEN RS485 AND RS232 WITHOUT CONVERTER Since a two wire RS485 output may be used directly on the RS 232 input of a PC or remo...

Страница 32: ...e directly connected to the load cell therefore calibrations and zero settings can be done The protocol used is Modbus the receiving instrument works as master and the transmitting one as slave Proced...

Страница 33: ...rial port set protocol set the serial communication parameters 2 RECEIVING INSTRUMENT see section SERIAL COMMUNICATION SETTING select the desired serial port set the mode It s not possible to enable t...

Страница 34: ...terminals of the line on terminal strip of the instrument furthest away in the network If should be there different instruments or converters refer to the specific manuals to determine whether it is n...

Страница 35: ...n Test It allows the analog signal to range between the minimum and the maximum values starting from the minimum current output test voltage output test Millivolt Test displays the load cell response...

Страница 36: ...act switching for decreasing weight For example with a setpoint at 100 and hysteresis at 10 the switching occurs at 90 for decreasing weight These values are set to zero if the calibration is changed...

Страница 37: ...nd time of Real Time Clock Confirm with ENTER to continue leave the instrument on for at least 12 hours to charge the battery if the alarm persists contact technical assistance an incorrect date has b...

Страница 38: ...PRINT key BASIC PRINTOUT W200 BASE Addr 01 DATE 12 09 11 14 48 12 GROSS 878 kg NET 589 kg TARE 289 kg BASIC PRINTOUT PEAK ENABLED W200 BASE Addr 01 DATE 12 09 11 14 48 12 GROSS 1204 kg NET 831 kg TARE...

Страница 39: ...xt is off to indicate that this menu item is unlocked TEMPORARY MENU UNLOCKING press and simultaneously for 3 seconds the unlock password is requested if enabled it is now possible to enter and modify...

Страница 40: ...or have not been set not approved program approved program single interval Dir 2014 31 EU art 1 approved program multi interval Dir 2014 31 EEU art 1 approved program multiple range Dir 2014 31 EU ar...

Страница 41: ...lock keypad lock if active when a key is pressed the message is displayed for 3 seconds keypad and display lock if active the kaypad is locked and the display shows the instrument model weight is not...

Страница 42: ...en bereinstimmt F D claration de conformit Nous d clarons avec cela responsabilit que le produit auquel se rapporte la pr sente d claration est conforme aux normes cit es ci apr s CZ Prohl en o shode...

Отзывы: