- 25 -
SERIAL COMMUNICATION SETTING
-
/
: communication port.
-
: it disables any type of communication (default).
-
: MODBUS-RTU protocol; possible addresses: from 1 to 99 (see Communication
protocols manual).
-
: ASCII bidirectional protocol; possible addresses: from 1 to 99 (see
Communication protocols manual).
-
-
-
: continuous weight transmission protocol (see Communication protocols
manual), at the frequency set in
item (from 10 to 300).
-
(set:
=
,
=
).
-
(set:
=
,
=
).
-
: continuous weight transmission protocol to RIP5/20/60, RIP50SHA, RIPLED series
remote displays; the remote display shows the net weight or gross weight according to its
settings (set:
=
,
=
,
=
).
-
: continuous weight transmission protocol to RIP6100, RIP675, RIP6125C series
remote displays; the remote display shows the net weight or gross weight according to its
settings (set:
=
,
=
,
=
).
-
: continuous weight transmission protocol to RIP6100, RIP675, RIP6125C series
remote displays (set:
=
,
=
,
=
).
When the remote display is set to gross weight:
-
if the instrument displays the gross weight, the remote display shows the gross
weight.
-
if the instrument shows the net weight, the remote display shows the net weight
alternated with the message
.
-
: printer.
-
: weight reception mode (see section
WEIGHT READING VIA SERIAL PORT
).
-
: weight reception mode (see section
WEIGHT READING VIA SERIAL PORT
).
-
: transmission speed (2400, 4800, 9600, 19200, 38400, 115200; default:
9600).
-
: instrument address (from 1 to 99; default: 1).
+