- 36 -
Upon instrument power-on hold down the ESC key until the display shows
, then proceed as
follows:
CONSTANTS RESTORE
(does not erase the calibration): confirm
, use arrow keys to select
, set code 6935 and confirm.
PROGRAM SELECTION:
confirm
and use the arrow keys to select the desired program:
: basic program, setpoint management only.
: to be used when the loaded weighing system correspond to not loaded cells and vice versa
(product increases while weight on load cells actually decreases).
: weight remote display program with setpoint.
Set the weight reading mode (except for
program):
-
: the weight is received either by load cells or intelligent junction boxes or transmitters
connected to the instrument.
-
: the weight is received via serial port (
WEIMOD
or
WEIRIP
mode).
Set the approval status (only if
or
have not been set)
-
: not approved program;
-
: approved program, single interval (Dir. 2014/31/EU, art. 1)*;
-
: approved program, multi-interval (Dir. 2014/31/EEU, art. 1)*;
-
: approved program, multiple range (Dir. 2014/31/EU, art. 1)*;
*
Contact technical assistance to request the proper manual and the correct procedures for
approval, indicating mandatory hardware code and serial number (see section
INSTRUMENT
COMMISSIONING
).
When a W series instrument is used in combination with an intelligent junction box or a
weight transmitter, the approval status set on both devices must be the same.
Configure the connection to the CLM serie intelligent junction box or to the weight transmitter
(only if
or
have not been set):
-
: intelligent junction box or transmitter connected to the instrument
-
: no intelligent junction box or transmitter connected
By confirming, the instrument is restored to default and data is erased.
If you do not have a specific manual for the newly set program, you can request it
to technical assistance.