- 29 -
-
: even parity
-
: odd parity
-
: stop bit (1 – 2; default: 1)
The transmitting instrument display is locked and shows the instrument model. To unlock
it, disconnect the receiving instrument and follow the procedure in section
KEYPAD OR
DISPLAY LOCKING
in the transmitting instrument manual.
WEIRIP MODE
The instrument receives the gross weight via serial port; calibrations and zero settings must be
performed on the transmitting instrument.
Procedure:
1.
TRANSMITTING INSTRUMENT
(see section
SERIAL COMMUNICATION SETTING
in the
transmitting instrument manual)
select the desired serial port
set
protocol
set the serial communication parameters
2.
RECEIVING
INSTRUMENT
(see section
SERIAL COMMUNICATION SETTING
)
select the desired serial port
set the
mode
It’s not possible to enable this function on several serial ports; in case of conflict,
the last serial set remains active.
set the serial communication parameters as on the transmitting instrument:
-
: transmission speed (2400, 4800, 9600, 19200, 38400, 115200; default: 9600)
-
: delay in milliseconds which elapses before the instrument replies (from 0 to
200 ms; default: 0)
-
:
-
: no parity (default)
-
: even parity
-
: odd parity
-
:
stop bit (1 – 2; default: 1)
set unit of measure (
) and number of decimals (
) of the gross weight received by
the transmitting instrument
The
and
menu items appear in the main menu after having set the
mode.