Annexes 

 

9 Annexes 

 

 

9.1 Water quality

 

 

In order to protect the water circuit of the ultracool units, the water to be 
cooled must have specific physical/chemical properties so that it is not 
aggressive. If this water is outside any of the limits listed in the table 
below, it can seriously damage some of the materials of the ultracool 
unit. 

 

Parameter Limit 

values 

pH 

7 – 8 

Total Hardness (TH) 

< 150 ppm 

Conductivity 

50 – 500 

µ

S/cm 

NH

3

 < 

ppm 

Total iron ions (Fe

2+

 and Fe

3+

< 0.2 ppm 

Chloride (Cl

-

< 300 ppm 

H

2

S < 

0.05 

ppm 

Solid particles 

< 150 

µ

 

The Total Hardness is specified in ppm (mg/L) of Ca

2

CO

3

 

Please note that ultra pure waters like deionised water can also be 
harmful for some of the materials of the ultracool units as they have a 
conductivity below 50 

µ

S/cm. 

 

The product refrifluid supplied by ultrafilter includes in its composition an 
anticorrosive, bactericide and antifreeze (until –7ºC / 19.4 ºF). It has the 
following characteristics: 

 

Parameter Limit 

values 

pH 7.5 

Total Hardness (TH) 

< 30 ppm 

Conductivity 

50 

µ

S/cm  

NH

3

 0 

ppm 

Total iron ions (Fe

2+

 and Fe

3+

< 0.2 ppm 

Chloride (Cl

-

) < 

ppm 

H

2

S 0 

ppm 

Solid particles 

< 50 

µ

 

A concentration of ethylene-glycol higher than the 30% can seriously 
damage the pump of the ultracool units.

 

 

ultrafilter will not accept any warranty for any damage caused by 
water that is out of one or more of the above limits. Therefore, if 
there is not a total security on the water quality we advise the use 
of our product refrifluid. This fluid has been specially designed for 
the protection of closed water circuits. Besides this ultrafilter offers 
the possibility of doing a basic analysis of a water sample supplied 
by the customer. 

 

UC-140 I

NDI

GO 

 50/

60Hz 29.

01.

03

 rev

. 0 

Содержание Ultracool Superplus series

Страница 1: ...OPERATION MANUAL ultracool mini 0140 Indigo Ref MI 080 Rev 1 Date 30 04 03 50 60 Hz...

Страница 2: ...The basic maintenance as indicated in point 5 1 should be carried out by properly trained personnel and if necessary under the supervision of a person qualified for this job ultrafilter personnel or p...

Страница 3: ...t up of the ultracool unit 7 3 3 Chiller strat up 8 4 Control Panel 4 1 Components of the control panel 9 4 2 Control thermostat 10 4 2 1 Operation 10 5 Maintenance 5 1 Basic maintenance 11 6 Troubles...

Страница 4: ...ded application in order to undertake additional safety measures if necessary 1 2 Safety regulations The operator has to observe the national working operating and safety regulations Also existing int...

Страница 5: ...compressor The ultracool unit must be transported by palet jack or forklift truck 2 3 Site The ultracool unit must be installed in an atmosphere where the range of temperatures is within the indicated...

Страница 6: ...nstallation the concentration of ethylene glycol should be checked Do not use automotive antifreeze Use lab grade ethylene glycol or refrifluid only Do not use an ethylene glycol concentration above t...

Страница 7: ...r outlet of the ultracool unit and a solenoid valve in the water inlet The supply of this solenoid valve will be carried out by terminals designed for that purpose see electrical diagrams 2 7 Electric...

Страница 8: ...of the ambient air Nominal 25 C 77 F Minimum 15 C 5 F 2 Maximum 50 C 122 F 1 For ambient temperatures below 5 C 40 F it is necessary to add ethylene glycol to the water of the circuit see point 2 4 3...

Страница 9: ...ial start up To do so start and stop the ultracool unit using the On Off switch it is easier to see the turn direction when the pump is stopping In the case of the pump rotating in the opposite direct...

Страница 10: ...when the compressor is working 7 Water level alarm indicator this is lit when the water level in the tank is too low It causes the ultracool unit to stop 8 Low flow temperature control indicator This...

Страница 11: ...probe In the 50Hz version the display shows the temperature in C and in the 60Hz version it shows it in F as depicted above Setting the temperature Follow these steps to introduce the required workin...

Страница 12: ...r filter if the pressure drop exceeds 0 5 bar 8 psi change the filter element Monthly With the Unit disconnected Main power switch Off clean the condenser with a blast of compressed air from the insid...

Страница 13: ...ation stopped Check the motor fan fuses If the problem persists contact authorised technical service Disconnect the chiller On Off and Main power switch both Off see point 4 Open the front panel of th...

Страница 14: ...following codes LO HI E0 Er2 EE The temperature has been below the setpoint for some minutes The temperature has been above the setpoint for some minutes Thermostat or temperature sensor is faulty If...

Страница 15: ...5 kW 1 1 1 0 max flow 4000 17 6 4000 17 6 min flow l h US gal min 400 1 8 400 1 8 max pres 6 3 91 6 3 91 Water pump min pres bar psi 3 3 48 3 3 48 Water tank volume litre US gal 100 26 100 26 Water co...

Страница 16: ...15 8 Log Book 8 Log Book 8 1 Log Book Date Remarks Signature UC 140 INDIGO 50 60Hz 29 01 03 rev 0...

Страница 17: ...below 50 S cm The product refrifluid supplied by ultrafilter includes in its composition an anticorrosive bactericide and antifreeze until 7 C 19 4 F It has the following characteristics Parameter Lim...

Страница 18: ...6003 E5326003 Sight glass 1 0107397000 0107397000 Refrigerant gas kg lb 14 31 R 134a R 134a Level control 1 E4202004 E4202004 Water by pass 1 E2301002 E2301002 Pump pressure gauge 1 E5544014 E5544022...

Страница 19: ......

Страница 20: ......

Страница 21: ......

Страница 22: ......

Страница 23: ......

Отзывы: