 Models  S 1200, S 2400, S 4400, POU 3300, POU 3500 

18

 - 

 

 

 

Open both the Fill and Service valves on the top of the module by pushing in and turning 

the valves counter-clockwise to the Open position.  

 

 

Press and hold the trigger lever on the canister to begin filling the reservoir.  

Continue to pressurize the canister, pumping slowly. 

 

 

Be watchful for process fluid flowing into the overflow tank (~ 500ml after the LL1 LED 

goes out @ 20 

º

C). Continue to fill until the fluid in the overflow tank is about 1 inch from 

its bottom. When no more air bubbles come out, release the trigger and stop pumping. 

The reservoir is full.  

 

 

Close both the Fill and Service valves by turning them clockwise to the Closed 

position. The module is now ready for system startup. 

 

3.2.

 

System Startup  

The system startup procedure requires that all connections between the  module and its PSC are 

connected and powered on, and all fluid connections are secured. Due to the small size of the  

unit’s tank, an additional fill procedure is needed to purge the entire process loop of air. 

 

 

Check that the EMO button (located on the right side of the controller front panel) is 

protruding. If pressed in, twist clockwise to reset. (The EMO Disable key is used during 

servicing only.)  

 

 

Turn on the PSC and verify that the PSC is displaying the temperature of fluid in the 

module. 

 

 

Press the START/STOP button to start the system, noting that the green LED (System On) 

on top of the unit should turn on. Depending on the amount of external plumbing, the 

red LED (LL1) will turn on to indicate a low tank volume. The pump continues to run if 

only LL1 is active. 

 

 

If the green system LED turns off, the fluid level is at LL2 and will shut down the system 

(unless in CHX mode)  

 

 

Open the Fill valve, and, using the fill canister, add process fluid. The red LED goes out 

when the reservoir is half filled. Continue to fill until the fluid flows into the overflow 

tank. Continue filling until the fluid level in the overflow tank is 2 inches from the bottom.  

 

 

Close the fill valve and disconnect the fill canister. Empty the excess process fluid into an 

authorized storage container. The system is now ready for temperature control 

operation. 

 

 

Содержание Semistat S 1200

Страница 1: ...Rev D 2 19 19 Temperature Control System MODELS Semistat S 1200 S 2400 S 4400 POU 3300 POU 3500 OPERATION MANUAL LAUDA Noah LP 2501 SE Columbia Way Ste 140 Vancouver WA 98661 USA...

Страница 2: ...eved by heating cooling a recirculating perfluorinated fluid and rejecting heat to the facility s water loop Each module is intended for installation with a LAUDA Noah Power Supply Controller PSC PROD...

Страница 3: ...ns_______________________________________________________ 15 3 Startup ________________________________________________________________ 17 3 1 Fill Procedure __________________________________________...

Страница 4: ...s Pin and signal names appear in capitalized italics DUTY CYCLE A Technical terms appear in italicized text when first introduced Unit labels switches indicators etc generally appear in boldface lette...

Страница 5: ...ity are within specifications before turning on the power supplies Use proper electrostatic discharge ESD precautions Do not attempt to operate product outside of its specified electrical range 1 3 Ri...

Страница 6: ...3500 5 Certification Certain options of this product are pending certification by CE certification EMC measurements SEMI S2 For more information refer to the letter of conformance US or declaration o...

Страница 7: ...stated directives and standards the following conditions of use must be met Operate the point of use module only within a system configuration that contains accessory components authorized and provide...

Страница 8: ...rrier Utilize the handles on the top of the unit for transport and use a 2 person lift for the POU 3500 and S 4400 2 3 Site and Climate The unit must be installed vertically according to the markings...

Страница 9: ...e denotes the climatic specifications of the unit Temperature Relative Humidity Air Pressure Operating 0 C to 50 C 32 F to 122 F 10 to 90 1 2 g m3 to 25g m3 1 Atm Storage and Transportation 20 C to 60...

Страница 10: ...Models S 1200 S 2400 S 4400 POU 3300 POU 3500 9 2 4 Identification of Ports and Connections The following illustration depicts the unit s external fittings and connectors...

Страница 11: ...nnector that can be adapted to either a barb fitting or a Swagelok compression tube adapter WARNING For applications where the temperature set point of the unit is less than the dew point of the envir...

Страница 12: ...4400 POU 3300 POU 3500 11 POU 3500 When connecting the process supply and return lines follow the following diagram Recall that the unit must be installed no more than 3 8m 12 5 ft away from the temp...

Страница 13: ...dup of moisture due to condensation Caution Working with Perfluorinated Fluids Contain any spills and absorb with available inorganic material Perfluorinated fluids are not considered hazardous but do...

Страница 14: ...ectionality of the water flow for optimal performance Using a flow meter verify that the unit is receiving the recommended flow rate for that particular module Be sure that shutoff valves for the inco...

Страница 15: ...Models S 1200 S 2400 S 4400 POU 3300 POU 3500 14 POU 3500...

Страница 16: ...ircular connector female end connecting with the module and a 25 pin D subminiature connector connecting to the PSC This is labeled Signal to Module Connect the PSC s AC power cable to either tool or...

Страница 17: ...econdary containment devices should not interfere with access to the PCW or drain fittings on the bottom of the unit and be designed for easy removal of the collected fluids Module Minimum volume of S...

Страница 18: ...not including external plumbing as well as a pressurized fill canister CAUTION Use only approved process fluids Do not use water or a mixture of water and ethylene glycol 3 1 Fill Procedure Fill the c...

Страница 19: ...ocedure is needed to purge the entire process loop of air Check that the EMO button located on the right side of the controller front panel is protruding If pressed in twist clockwise to reset The EMO...

Страница 20: ...EMOTE LOCAL Switches the system s operation mode from tool controlled REMOTE to front panel controlled LOCAL CHANGE SETPOINT Press this button to initiate a change in the process setpoint while in loc...

Страница 21: ...move the unit the unit must be detached from the facility connections water electricity and CDA 1 Identify the connections to CDA Facility water and electrical connectors to the PSCX a CDA is connecte...

Страница 22: ...rocedure Press the Start Stop key to enter idle mode then toggle the power switch off Depressurize the fill canister by unscrewing the trigger assembly from the top of the canister Leave the bottom ho...

Страница 23: ...ally FAN FLOW FAULT Internal fan is not functioning properly Contact LAUDA for service when able Hard Alarm Display Description Solution SEMISTAT POWER SUPPLY INTERRUPTION Power supply for thermoelect...

Страница 24: ...elow this value the power supply outputs pump thermoelectrics heater are turned off Verify that the process value limits on the PSC are set to values within the range of the application TE OVERTEMP Op...

Страница 25: ...he EMO OUT input is not used it must be jumpered Connector is a 9 pin D subminiature connector labeled EMO OUT on the back of the controller Connector is supplied with the unit Process temperature flu...

Страница 26: ...leaks If leakage problems persist try using two continuous hoses between the module and the process equipment Controller does not communicate with host equipment 1 Controller is not in Remote mode 2...

Страница 27: ...1400 1500 2600 Cooling capacity at 10 C W 900 950 1925 2000 3500 Cooling capacity at 20 C W 1200 1200 2450 2400 4400 Heater Heating capacity W 3000 3000 6000 6000 12000 Temperature Working temperature...

Страница 28: ...33 1 25 66 2 50 66 2 50 Interfaces Barb Process Barb Process Barb Process Barb Process Barb Process Swagelok Compression PCW Swagelok Compression PCW Swagelok Compression PCW Swagelok Compression PCW...

Страница 29: ...120 2 3 Dewpoint C 30 or lower 30 or lower 30 or lower 30 or lower 30 or lower User Interface Display N A N A N A N A N A Value input PSC Keypad PSC Keypad PSC Keypad PSC Keypad PSC Keypad Safety Con...

Страница 30: ...0 5 to 40 5 to 40 Relative humidity maximal 80 at 31 C maximal 80 at 31 C maximal 80 at 31 C maximal 80 at 31 C maximal 80 at 31 C Operational area for internal use altitude up to 2000 m for internal...