Proline Kryomats
53 / 137
7.5.4
Activate external pump
As an option an external pump is available for the Proline Kryomats for external applications. This pump can be set on /
off manually in the menu shown below. Another possibility is the automatic mode. In this case the pump is switched
according to the unit status standby / running.
Command
Ext. Pump
Open the device parameter menu via the soft
key Menu .
Change from
Pump
Ext. pump
using .
With
or you enter the illustrated
window.
Use
or
to switch the pump off,
permanently on or to set the automatic mode.
Confirm your selection with
or End ,
or quit the window with without making any
changes.
Automatic mode means that the external pump
is switched off while the unit is in standby mode
or it is switched on while the unit is running.
7.5.5
Activating external control
An external temperature probe can be connected to the Proline Thermostats. How this is done is explained in Section
7.5.2. If the bath temperature is to be controlled using this sensor instead of the internal sensor, the setting can be
made here.
Furthermore, control can also occur based on the signal from the analog or serial module (
4.8).
Master
Con
and 2 x
Call the source selection for the control
Con
.
The momentary setting for the source is displayed,
here
1
for internal, because control takes place using the temperature signal
from the internal temperature probe.
The source display flashes.
Con 1
°C
Pump Menu
End
T
set
T
fix
off
on
auto.
Ext. pump
Содержание RP 3090 C
Страница 130: ...130 137 Proline Kryomats 03 2022 12 Declaration of conformity and product returns declaration ...
Страница 131: ...03 2022 Proline Kryomats 131 137 ...
Страница 134: ...134 137 Proline Kryomats 03 2022 ...
Страница 135: ......