Proline Kryomats
63 / 137
The temperature measurement device shows
the true temperature value (with glass
thermometers consider the correction where
applicable!).
Change the display in the adjacent window to
the true value with cursor or soft keys and
accept with
or End ,
or quit the window with without making
changes.
7.6.7
Restoring the works setting of the internal temperature-probe offset
If the offset has been misadjusted unintentionally, the works setting can be restored with this function.
Command
Default
Open the device parameter menu via the soft
key Menu .
With the cursor keys continue to
Settings
Calibration
intern Pt100
Default
.
The window shown adjacent appears.
Confirm selection with
.
Confirm the control dialog on the right with
or cancel with .
Return to the measurement window
with End or .
Help
Menu Screen
T
set
T
fix
6
7
8
9
0
1
2
3
4
5
Temperature value of the ref.
temp. measurement device:
20,
15
Min: -50.0°C Max: 302.00
Pump Menu
End
T
set
T
fix
intern Pt100
extern Pt100
Calibration
Default
Pump Menu
End
T
set
T
fix
Confirm input!
Enter key: Continue
Escape key: Cancel