Proline Kryomats
61 / 137
7.6.4
Restoring works settings
All works settings, apart from the control parameters and the sensor calibration, are restored.
Command
Works settings
Open the device parameter menu via the soft
key Menu .
With the cursor keys continue to
Settings
Works settings
.
The window shown opposite appears.
Master and then
only control par. int.
is
shown as a possible choice.
There are however various possibilities, which
can be selected with
or
:
Under
All modules
Master, Command and all
connected modules are reset to the works
setting with
all default
.
Under
Master
you have the choice between:
all default
, then all Master settings are
reset,
only control para. int.
for the internal
control parameters,
only control para. ext.
similar for
external,
only miscellaneous
which resets setpoint,
pump level, maximum current
consumption, control to internal and auto
start to "Auto".
Under
Command
all command settings are
reset with
All default
.
Confirm selection with
.
Confirm the control dialog shown on the left
with
or cancel with .
Return to measurement window with
End or .
Pump Menu
End
T
set
T
fix
All modules
Master
Command
Cool
all default
only control par. int
only control par. ext
only miscellaneous
Pump Menu
End
T
set
T
fix
Confirm input!
Enter key: Continue
Escape key: Cancel