Hose type
Length in cm
Temperature range of
the hose in °C
Application area
Catalog number
MK 150
150
-90 – 150
With foam insulation for the refrig-
eration sector, suitable for all
LAUDA heat transfer liquids
LZM 054
MK 200
200
-90 – 150
With foam insulation for the refrig-
eration sector, suitable for all
LAUDA heat transfer liquids
LZM 055
Pump short-circuit
18
-90 – 150
With foam insulation for the refrig-
eration sector, suitable for all
LAUDA heat transfer liquids
LZM 045
4.8.2 Connecting an external consuming unit
DANGER!
Risk of hot heat transfer liquid overflowing
Fire
A connecting hose with a collecting vessel must be
attached to the overflow.
The collecting vessel and connecting hose must be designed
to withstand the maximum temperature of the heat transfer
liquid.
Avoid ignition sources in the vicinity of the collecting vessel.
WARNING!
Degradation of overflow or drain
Electric shock
Overflow hose and drainage hose must each slope down-
ward separately all the way into a collecting vessel.
CAUTION!
Risk of external hydraulic circuit bursting from overpressure
Scalding, cold burns
When laying the hoses, make sure they cannot kink.
Use safety valves in the hydraulic circuit.
CAUTION!
Risk of heat transfer liquid escaping during operation due to
open consuming unit
Scalding, cold burns
Always use hydraulically sealed consuming units.
V6
PRO bath thermostats and circulation thermostats
54 / 156