The [Level] tile indicates the bath level on a scale from 9 to 0.
1.
Turn on the device and the pump. Set the bath temperature to room
temperature.
2.
Reduce the bath level by draining heat transfer liquid from the
draining nozzle.
The [Level] tile displays the drop of the heat transfer liquid on the
screen.
If the liquid falls below level 1, the device switches off and a
window containing the message
Alarm
appears on the screen.
The visual operation and fault indicator (LED) flashes red.
3.
Press [OK] to confirm the message on the screen.
The window disappears from the screen.
4.
Top up the heat transfer liquid.
The [Level] tile displays the liquid level on the screen.
5.
Press the red button to unlock. This button is located on the back or
side of the device, depending on the device.
The red LED goes out. The signal tone is turned off.
V6
PRO bath thermostats and circulation thermostats
119 / 156
Содержание P 10
Страница 146: ...V6 PRO bath thermostats and circulation thermostats 146 156 ...
Страница 147: ...V6 PRO bath thermostats and circulation thermostats 147 156 ...
Страница 148: ...V6 PRO bath thermostats and circulation thermostats 148 156 ...
Страница 157: ......
Страница 158: ......
Страница 159: ......