5.
Open the drain tap by turning it counterclockwise.
Drain the bath, external consuming unit, accessories and
hoses completely.
6.
If necessary, clean or flush out the device (with new heat transfer
liquid, for example).
After changing to a different heat transfer liquid, you may have to
set new values for the temperature limit, overtemperature switch-off
point and/or the controller output limit.
5.4 Establishing a mains connection
WARNING!
Contact with voltage conductors due to faulty mains cable
Electric shock
The mains cable must not come into contact with the top
of the device, hoses containing hot heat transfer liquid
and other hot parts, neither during operation nor after the
device is switched off.
NOTICE!
Use of impermissible mains voltage or mains frequency
Device damage
Compare the type plate with the available mains voltage and
mains frequency.
Please note the following:
Note for electric installation on site:
The devices must be protected with a 16 ampere circuit breaker
fitted during installation.
Exception: Devices with 13 ampere UK plugs.
Only use the supplied power cable for the power supply.
Only connect the device to sockets with a protective earth conductor
(PE).
V6
PRO bath thermostats and circulation thermostats
65 / 156