5.
Switch on the device at the mains switch.
A signal tone is emitted. The software version is shown in the
display. The actual temperature is then shown in the display.
The constant temperature equipment starts operation, the pump
is started.
Depending on the set temperature, the cooling unit is started
after 2 minutes at the earliest. The blue LED is lit if the cooling
unit is active.
6.
Switch on the device at the mains switch.
A signal tone is emitted. The software version is shown in the
display.
FILL
is then shown in the display. The pump and the
cooling unit are not started if the level is low.
Fill the device with heat transfer liquid.
7.
Carefully pull the cover on the filler nozzle upwards, do not twist it.
8.
Carefully pour heat transfer liquid into the filler nozzle. Monitor the
level indication. Fill the device to the maximum fill level.
Use a funnel for filling if required.
The level indication must not be above the maximum fill level.
9.
Press the Enter key when the maximum or a sufficient fill level is
reached.
The pump starts. The actual temperature is shown in the display.
The fill level falls as the consuming unit is filled.
10. Carefully top up the heat transfer liquid, as the consuming unit is
now being filled. If the fill level falls too low, the device switches
automatically to FILL mode and the pump and the cooling unit are
switched off. Continue filling until trouble-free operation is possible.
Monitor the level indication.
11.
Carefully press the cover in the filler nozzle.
5.4 Setting the pump pressure
The pump pressure can be adjusted via a control valve on the rear of the
device in devices with a bypass (MC 600 and MC 1200). The pump
pressure can therefore be set individually for pressure-sensitive external
consuming units.
Open the bypass adjusting wheel on the back of the device to the
full extent before switching the device on.
Turn the adjusting wheel counterclockwise to open.
Fill level sufficient
Fill level too low (low level)
Fill level falls
V07
Microcool
29 / 57