SoundAdvisor Model 831C
Setting Up a Measurement With Measurement History
15-1
Module 15
Using Measurement History
T
his module shows how to use the Measurement History feature of
firmware option 831C-ELA, which allows you to segment the data gathered
between Run and Stop into measurement intervals. Measurement intervals
are saved as individual records in a single data file.
Option 831C-ELA provides exceedance-based logging analysis with manual or
timed storage of statistical data and measurement history.
In this module:
15.1 Setting Up a Measurement With Measurement History
Using Measurement History when the measurement starts, the meter
automatically creates a series of measurement intervals. You define the
duration of each interval. When the measurement stops, all intervals are
saved together in one file.
While measurement parameters can be indicated directly in the Active setup,
it’s a good practice to save a setup file for each specific measurement you
perform as shown in this section.
TAKE NOTE
The G4 LD Utility also provides a simple, file transfer-style interface for
completing this same process. For more information, refer to the
G4 LD Utility
Manual
.
Before you begin:
•
Verify that firmware option 831C-ELA is installed and enabled on your
meter. For more information, see
.
•
Create a measurement setup file (Setup) as shown in
Custom Setup File On the Meter
.
Step 1
Go to
Tools
3
Setup Manager.
Step 2
Highlight the name of the Setup file you want to configure with Measurement
History.
Step 3
Press the top-right softkey (
//*
) until the
Control
tab appears.
Step 4
Select a
Run Mode
and associated settings to complete the Setup. Click a link
below for instruction on completing the Setup:
•