Accuracy Grade
Model 831 Manual
Cu
– Cu is the degree of curvature, and is a comparison of
the T20 and T30. Cu is expressed as the percentage
deviation from being perfectly in-line. See ISO 3382-
2:2008(E), Annex B.3.
SD
– SD is the standard deviation of the measurement
results for the T30 or the T20 decay times. See equations 2
and 3 in ISO 3382-2:2008(E), Sections 7.1. For impulsive
excitation, n = 10 is used as defined in section 7.2
Accuracy Grade
ISO 3382-2:2008(E) describes three methods of differing
measurement uncertainty as follows:
Survey Method
The survey method is appropriate for the assessment of the
amount of sound absorption for noise control purposes, and
survey measurements of the airborne and impact sound
insulation. It should be used for measurements in ISO
10052. Survey measurements are made in octave bands only.
The nominal accuracy is assumed to be better than 10% for
octave bands.
Make measurements of the reverberation time for at least
one source position. Find the average of results from at least
two source-microphone combinations, see Table 15 - 3.
Engineering Method
The engineering method is appropriate for verification of
building performance for comparison with specification of
reverberation time or room absorption. It should be used for
measurements in ISO 140 (all parts) with remarks to
reverberation time measurements. The nominal accuracy is
assumed to be better than 5% in octave bands and better than
10% in one-third octave bands. See Table 15 - 3.
Measure reverberation time two or more times for each
source-microphone combination. At least six independent
source-microphone combinations are required, see Table 15
- 3 'Minimum Number of Positions and Measurements'.
Precision Method
The precision method is appropriate where high
measurement accuracy is required. The nominal accuracy is
assumed to be better than 2.5% in octave bands and better
than 5% in one-third-octave bands.
Содержание 831
Страница 1: ...Model 831 Sound Level Meter Manual...
Страница 2: ...I831 01 Rev O Manual Larson Davis Model 831 Manual...
Страница 123: ...5 44 Any Level Display Model 831 Manual...
Страница 135: ...6 12 Run Modes Without Measurement History Model 831 Manual...
Страница 155: ...7 20 Recovery After Improper Shutdown Model 831 Manual...
Страница 209: ...11 22 Markers Model 831 Manual...
Страница 218: ...Model 831 Manual Display of Measurement History Data 12 9 FIGURE 12 7 Measurement History Displays...
Страница 221: ...12 12 Link to Time History Model 831 Manual FIGURE 12 11 Time History Display...
Страница 277: ...15 10 Viewing and Analyzing Results Model 831 Manual FIGURE 15 15 Navigation RIGHT 5 ENTER 5 ENTER...
Страница 321: ...16 24 Sound Recording Playback Model 831 Manual...
Страница 356: ...Model 831 Manual Displays 18 25 FIGURE 18 38 Live Tab Displays FIGURE 18 39 Overall Tab Displays...
Страница 357: ...18 26 Displays Model 831 Manual Session Log Display FIGURE 18 40 Session Log Displays FIGURE 18 41 Current Displays...
Страница 358: ...Model 831 Manual Displays 18 27 FIGURE 18 42 Measurement Displays FIGURE 18 43 Events Displays...
Страница 393: ...20 18 Status Tab Model 831 Manual RS 232 FIGURE 20 21 Status RS 232 USB FIGURE 20 22 Status USB Channel...
Страница 403: ...20 28 Troubleshooting Model 831 Manual...
Страница 413: ...21 10 Calibration When The Model 831 Is Locked Model 831 Manual...
Страница 421: ...23 4 File System Model 831 Manual...
Страница 501: ...A 64 Vibration Sensitivity Model 831 Manual...
Страница 513: ...B 12 Sections 5 6 7 and 9 except 9 3 Model 831 Manual...
Страница 525: ...C 12 Community Noise Descriptors Model 831 Manual...