Model 831 Manual
1/1 and 1/3 Octave Filters
A-23
1/1 Octave Filter Linearity
The filter linearity range depends upon both the Input Gain, which is set as described in the section "20
dB Gain" on page 4-5, and the Filter Range, which is set as described in the section "OBA Range
Setting" on page 4-7.The data, which were measured using an electrical input, are in dB re. 1
V,
which is equivalent to dBSPL when a 50 mV/Pa microphone is used
Table A-20 1/1 Octave Linearity Range
1/1 Octave Filter Summation Error
The Octave Filter Summation Error shown in the following graphs is defined by IEC61260 4.9.
FIGURE A-4 Gain = 0 dB, Filter Range = Normal
Gain
0 dB
0 dB
20 dB
20 dB
Range
Normal
Low
Normal
Low
Frequency
Overload
Linearity
Range
Lower
Linearity
Limit
Noise
Floor
Overload
Linearity
Range
Lower
Linearity
Limit
Noise
Floor
Overload
Linearity
Range
Lower
Linearity
Limit
Noise
Floor
Overload
Linearity
Range
Lower
Linearity
Limit
Noise
Floor
8.0
140.8
78.9
61.9 31.9
107.5
75.5
32.0
14.1
120.7
83.7
37.0
16.2
87.4
52.4
35.0
13.7
16.0
140.8
85.8
55.0 31.9
107.5
78.7
28.8
12.8
120.7
88.7
32.0
14.7
87.4
59.4
28.0
10.0
31.5
140.8
87.8
53.0 30.4
107.5
80.5
27.0
8.8
120.7
89.7
31.0
14.3
87.4
58.4
29.0
9.0
63.0
140.8
92.8
48.0 29.5
107.5
81.5
26.0
6.2
120.7
90.7
30.0
12.9
87.4
60.4
27.0
5.8
125.0
140.8
95.8
45.0 31.1
107.5
84.5
23.0
4.4
120.7
89.7
31.0
12.6
87.4
65.4
22.0
4.1
250.0
140.8
94.8
46.0 31.1
107.5
90.5
17.0
2.2
120.7
92.7
28.0
11.2
87.4
70.4
17.0
2.8
500.0
140.8
90.8
50.0 33.5
107.5
92.5
15.0
2.1
120.7
91.7
29.0
12.1
87.4
69.4
18.0
2.0
1000.0
140.8
95.7
45.1 31.7
107.5
88.5
19.0
2.0
120.7
95.7
25.0
12.0
87.4
70.4
17.0
0.2
2000.0
140.8
92.7
48.1 33.6
107.5
92.5
15.0
2.5
120.7
95.6
25.1
13.8
87.4
75.4
12.0
-2.9
4000.0
140.8
90.8
50.0 36.7
107.5
90.5
17.0
5.1
120.7
92.6
28.1
16.7
87.4
77.4
10.0
-2.3
8000.0
140.8
88.8
52.0 40.3
107.5
87.5
20.0
8.3
120.7
87.7
33.0
20.2
87.4
75.4
12.0
-0.4
16000.0
140.8
85.8
55.0 43.6
107.5
84.5
23.0
11.5
120.7
85.7
35.0
23.5
87.4
73.4
14.0
2.0
1/1 Octave Sum m ation
(UURU
-2
-1.5
-1
-0.5
0
0.5
1
1.5
2
8.058 18.84 44.03 102.9 240.6 562.3 1314 3073 7182
Frequency
S
u
mma
ti
o
n
(UURU
(d
B
)
Содержание 831
Страница 1: ...Model 831 Sound Level Meter Manual...
Страница 2: ...I831 01 Rev O Manual Larson Davis Model 831 Manual...
Страница 123: ...5 44 Any Level Display Model 831 Manual...
Страница 135: ...6 12 Run Modes Without Measurement History Model 831 Manual...
Страница 155: ...7 20 Recovery After Improper Shutdown Model 831 Manual...
Страница 209: ...11 22 Markers Model 831 Manual...
Страница 218: ...Model 831 Manual Display of Measurement History Data 12 9 FIGURE 12 7 Measurement History Displays...
Страница 221: ...12 12 Link to Time History Model 831 Manual FIGURE 12 11 Time History Display...
Страница 277: ...15 10 Viewing and Analyzing Results Model 831 Manual FIGURE 15 15 Navigation RIGHT 5 ENTER 5 ENTER...
Страница 321: ...16 24 Sound Recording Playback Model 831 Manual...
Страница 356: ...Model 831 Manual Displays 18 25 FIGURE 18 38 Live Tab Displays FIGURE 18 39 Overall Tab Displays...
Страница 357: ...18 26 Displays Model 831 Manual Session Log Display FIGURE 18 40 Session Log Displays FIGURE 18 41 Current Displays...
Страница 358: ...Model 831 Manual Displays 18 27 FIGURE 18 42 Measurement Displays FIGURE 18 43 Events Displays...
Страница 393: ...20 18 Status Tab Model 831 Manual RS 232 FIGURE 20 21 Status RS 232 USB FIGURE 20 22 Status USB Channel...
Страница 403: ...20 28 Troubleshooting Model 831 Manual...
Страница 413: ...21 10 Calibration When The Model 831 Is Locked Model 831 Manual...
Страница 421: ...23 4 File System Model 831 Manual...
Страница 501: ...A 64 Vibration Sensitivity Model 831 Manual...
Страница 513: ...B 12 Sections 5 6 7 and 9 except 9 3 Model 831 Manual...
Страница 525: ...C 12 Community Noise Descriptors Model 831 Manual...