HAPTER
5:
Advanced Operation
Model 336 Temperature Controller
Setting resolution is also six digits in sensor units. The curve format parameter
defines the range and resolution in sensor units as shown in TABLE 5-3. The sensor
type determines the practical setting resolution. TABLE 5-4 lists recommended sen-
sor units resolutions.
Enter the breakpoints with the sensor units value increasing as point number
increases. There should not be any breakpoint locations left blank in the middle of a
curve. The search routine in the Model 336 interprets a blank breakpoint as the end of
the curve.
5.9 Front Panel
Curve Entry
Operations
There are five operations associated with front panel curve entry: Edit curve, View
Curve, Erase Curve, Copy Curve, and SoftCal; as detailed below.
Menu Navigation:
Curve Entry
Q
(
Edit Curve, View Curve, Erase Curve, Copy Curve, SoftCal
)
5.9.1 Edit Curve
Use the Edit Curve operation to enter a new curve or edit an existing user curve. Only
user curves (21 to 59) can be edited. Entering the identification parameters associ-
ated with the curve is as important as entering the breakpoints. Curve header param-
eters are listed in TABLE 5-3. Typical curve parameters for common sensors are listed
in TABLE 5-4. Read this section completely and gather all necessary data before
beginning the process.
If the curve you wish to enter has similar parameters as an existing curve, first copy the
similar curve (as described in Section 5.2.4) to a new location, then edit the curve to the
desired parameters.
Sensor Type
Typical Lake Shore
Model
Format
Limit (K)
Temperature
Coefficient
Typical Sensor
Resolution
Silicon Diode
DT-670
V/K
475
Negative
0.00001 V
GaAlAs Diode
TG-120
V/K
325
Negative
0.00001 V
Platinum 100
PT-100
)
/K
800
Positive
0.001
)
Platinum 1000
–*
)
/K
800
Positive
0.01
)
Rhodium-Iron
RF-800
)
/K
325
Positive
0.001
)
Carbon-Glass
CGR-1-1000
log
)
/K
325
Negative
0.00001 log
)
Cernox™
CX-1050
log
)
/K
325
Negative
0.00001 log
)
Germanium
GR-200A-100
log
)
/K
325
Negative
0.00001 log
)
Rox™
RX-102A
log
)
/K
40
Negative
0.00001 log
)
Type K
9006-005
mV/K
1500
Positive
0.0001 mV
Type E
9006-003
mV/K
930
Positive
0.0001 mV
Type T
9006-007
mV/K
673
Positive
0.0001 mV
Au-Fe 0.03%
–*
mV/K
500
Positive
0.0001 mV
Au-Fe 0.07%
9006-001
mV/K
610
Positive
0.0001 (mV)
*Not offered by Lake Shore
TABLE 5-4
Typical curve parameters
Operation
Description
Refer to section:
Edit Curve
Edit Curve allows you to edit curves at any user curve location. Stan-
dard curves cannot be changed.
View Curve
View Curve allows you to view any curve at any curve location. No
curves can be changed.
Erase Curve
Erase Curve allows you to delete a curve from any user curve location.
Standard curves cannot be erased.
Copy Curve
Copy Curve allows you to copy a curve from any location to any user
curve location. Curves cannot be copied into standard curve locations.
SoftCal
SoftCal allows you to create a new temperature curve from a standard
curve and known data points entered by the user.
TABLE 5-5
Front panel curve entry operations
Содержание 336
Страница 4: ...Model 336 Temperature Controller...
Страница 6: ...Model 336 Temperature Controller...
Страница 26: ...14 cHAPTER 1 Introduction Model 336 Temperature Controller...
Страница 54: ...42 cHAPTER 3 Installation Model 336 Temperature Controller...
Страница 84: ...72 cHAPTER 4 Operation Model 336 Temperature Controller...
Страница 104: ...92 cHAPTER 5 Advanced Operation Model 336 Temperature Controller...
Страница 164: ...152 cHAPTER 7 Options and Accessories Model 336 Temperature Controller...
Страница 178: ...166 cHAPTER 8 Service Model 336 Temperature Controller...